- 5 -
Pump bracket ensures pump remains securely in position.
iv.
Electrical
The voltage of the Ultrapol Advance unit is switch able 110/220V unless otherwise stated by means of
a label on the machine. Select the proper input voltage via the selector switch on the back of the base
unit. The electrical power cord should be attached to the machine, and then the opposite end
connected to the 110/220V wall socket. Power for the Mast head assembly is achieved by connection
of the black wire / connector extending from the tool head into the labeled socket on the back of the
Base Unit. Place the black switch, which is right next to the power cord socket on the back of the Base
Unit, to the ON position.
Place bracket over
basin and tighten
screw to fix
position. Place
pump directly
over the bracket as
shown in the
image to the right.
Drain
Power for
Recirculation
Pump
Water
Supply Inlet
Main Power
Receptacle
Voltage
Selector
110V Outlets
(Only plug Ultracollimator