2. General Safety
The Kronos flash photolysis spectrometer is designed and manufactured for kinetic analysis of
chemical specimen by means of ultrafast transient absorption spectrometry. And it is not sold, nor
intended for, nor should ever be used for any other purpose. The product should be used solely in
accordance with the instructions provided.
2.1 Hazards
Take extra caution when operating or servicing this equipment:
•
If this equipment is used in a manner not specified in this manual, the protection provided by
this equipment may be impaired.
•
Handle the cuvette with care. Check the cuvette for damage or cracks and replace any damaged
ones immediately.
•
Use extreme caution in handling the cuvette when it is filled with liquid. Always remove the
cuvette before changing contents. Never fill or refill the cuvette when it is in the sample holder.
•
The optimal operation temperature of Kronos is at or near room temperature.
2.2 Electrical Safety
Observe these general warnings when operating or servicing this equipment:
•
Read all warnings on the unit and in the operating instructions.
•
Kronos uses a large capacitor to charge the Xe flash lamp. Dangerous voltage levels are
expected to persist even after unplugging the unit.
•
Do not disassemble or open the spectrometer case. Doing so may damage the excitation light
source and risk electric shock.
•
Do not use this equipment in or near water.
•
Dry the exterior of the cuvette prior to placing it in the holder. Using a cuvette with a wet exterior
or one that is leaking liquid can result in malfunction, shorting of electrical components or
electric shock.
6