SonixMDP/SP/OP Service Manual 00.053.205, Revision A
Chapter 7: Field Service Components
7-7
7.1.5
Backing up/Exporting Patient Data via the Service Mode Option
This function allows Service Technicians to backup (and restore)
Patient/Exam
data. It does not, however, backup
any of the
User Data
settings (
).
Patient/Exam
data will be copied to an auto-created directory at the root of the USB device called PatientInfo. If
this directory – and therefore patient data – already exists, the system will warn the user, allowing them to exit the
backup procedure without overwriting existing data. Choosing to overwrite the data should be done with the utmost
caution.
For details on restoring data via this option, refer to
Equipment/Tools Required:
• USB keyboard
• USB media (memory stick, external hard drive, etc.) with at least 5 Gb of available space.
To Backup/Export Patient Data (via the Service Mode Option):
1.
Connect the USB keyboard to the one of the four USB ports.
2.
Connect the USB storage device to the one of the three remaining USB ports.
3.
Press the console
button.
4.
Select
Administrator > Service…
.
5.
Type in the
Service Mode Password
using the USB keyboard.
6.
Select
OK
.
7.
From the
Advanced Tools and Options
drop-down menu, select
Patient Data Backup/Restore
.
Note:
If
Patient/Exam
data is backed up using these instructions (
), it will not be necessary to use the
Export
option (
Cautions:
The
Backup
function will overwrite any
Patient/Exam
data that already exists on the USB device.
Due to privacy regulations and laws, Service Technicians are forbidden from removing patient data from any client
site. Refer to
for more details.
Note:
USB storage space required to backup
Patient/Exam
data depends on the number of patients/images on
the system. Hard drive capacity is 160 Gb in total.
Summary of Contents for Sonix MDP
Page 1: ...Analogic Ultrasound Service Manual SonixMDP SP OPQ Ultrasound System...
Page 2: ......
Page 4: ......
Page 14: ...Chapter 2 System Specifications 00 053 205 Revision A SonixMDP SP OP Service Manual 2 2...
Page 46: ...Chapter 5 Software 00 053 205 Revision A SonixMDP SP OP Service Manual 5 4...
Page 120: ...Chapter 8 DICOM 00 053 205 Revision A SonixMDP SP OP Service Manual 8 24...
Page 147: ......