Appendix A: Troubleshooting Solutions
00.053.101, Revision C SonixMDP/SP/OP Service Manual
A-12
To Configure the BIOS:
1.
Ensure a USB keyboard is connected to the system.
2.
Access the BIOS.
3.
Select
BIOS > Integrated Peripherals > Gbe Controller
.
4.
Select
Disabled
.
To Physically Check the Network Cable Connection:
1.
Access the modulo (
2.
Ensure the internal network cable is connected securely to the
second
network port on the System Case
Connectivity Panel.
To Configure a Static IP Address:
1.
Ensure a USB keyboard is connected to the system.
2.
Press the
WINDOWS START KEY
. and select
Settings > Control Panel > Network Connections > Local
Area Connection
.
3.
Select
Properties > General > Internet Protocol (TCP/IP)
.
4.
Select
Properties
.
5.
Select
Use the following IP address
and enter the relevant information.
6.
Select
OK
and exit.
A.14 TRANSDUCER CONNECTION PORT NOT FUNCTIONING
To Confirm MUX Board Needs Replacing:
1.
Disconnect all transducers.
2.
Reconnect only one (1) transducer.
3.
Restart the system.
4.
If the transducer still isn’t functioning, the MUX board needs to be replaced (Contact Ultrasonix Technical
Support for the latest release of
Service Note 0126
, UXID 00.063.121).
Note:
Intel Controller
should be
Enabled
.
Note:
for details on the location of the network port.
Note:
Some customers may require a
Static IP Address
. If this is the case, use the following configuration
directions.
Note:
It may be necessary to restart in order for the changes to take affect.
Summary of Contents for SonixMDP
Page 1: ...SonixMDP SP OP Ultrasound System Service Manual...
Page 2: ......
Page 4: ......
Page 10: ...Table of Contents 00 053 101 Revision C SonixMDP SP OP Service Manual vi...
Page 12: ...List of Service Drawings 00 053 101 Revision C SonixMDP SP OP Service Manual ii...
Page 44: ...Chapter 4 Performance Testing 00 053 101 Revision C SonixMDP SP OP Service Manual 4 4...
Page 48: ...Chapter 5 Software 00 053 101 Revision C SonixMDP SP OP Service Manual 5 4...
Page 124: ...Chapter 9 Network Configuration 00 053 101 Revision C SonixMDP SP OP Service Manual 9 8...
Page 142: ...Chapter 11 Maintenance 00 053 101 Revision C SonixMDP SP OP Service Manual 11 16...
Page 150: ...Chapter 12 Troubleshooting Issues 00 053 101 Revision C SonixMDP SP OP Service Manual 12 8...
Page 166: ...Appendix C Ultrasonix Limited Warranty 00 053 101 Revision C SonixMDP SP OP Service Manual C 2...
Page 168: ...Appendix D Service Drawings 00 053 101 Revision C SonixMDP SP OP Service Manual D 2...