SonixMDP/SP/OP Service Manual 00.053.101, Revision C
Chapter 7: Field Service Components
7-7
7.1.5
Backing up/Exporting Patient Data via the Service Mode Option
This function allows Service Technicians to backup (and restore)
Patient/Exam
data. It does not, however, backup
any of the
User Data
settings (
).
Patient/Exam
data will be copied to an auto-created directory at the root of the USB device called PatientInfo. If
this directory – and therefore patient data – already exists, the system will warn the user, allowing them to exit the
backup procedure without overwriting existing data. Choosing to overwrite the data should be done with the utmost
caution.
For details on restoring data via this option, refer to
Equipment/Tools Required:
• USB keyboard
• USB media (memory stick, external hard drive, etc.) with at least 5 Gb of available space.
To Backup/Export Patient Data (via the Service Mode Option):
1.
Connect the USB keyboard to the one of the four (4) USB ports.
2.
Connect the USB storage device to the one of the three (3) remaining USB ports.
3.
Press the console
button.
4.
Select
Administrator > Service…
.
5.
Type in the
Service Mode Password
using the USB keyboard.
6.
Select
OK
.
Note:
If
Patient/Exam
data is backed up using these instructions (
), it will not be necessary to use the
Export
option (
Cautions:
The
Backup
function will overwrite any
Patient/Exam
data that already exists on the USB device.
Due to privacy regulations and laws, Service Technicians are forbidden from removing patient data from any client
site. Refer to
and
7.6.1 Removing the Hard Drives
for more details.
Note:
USB storage space required to backup
Patient/Exam
data depends on the number of patients/images on
the system. Hard drive capacity is 160 Gb in total.
Summary of Contents for SonixMDP
Page 1: ...SonixMDP SP OP Ultrasound System Service Manual...
Page 2: ......
Page 4: ......
Page 10: ...Table of Contents 00 053 101 Revision C SonixMDP SP OP Service Manual vi...
Page 12: ...List of Service Drawings 00 053 101 Revision C SonixMDP SP OP Service Manual ii...
Page 44: ...Chapter 4 Performance Testing 00 053 101 Revision C SonixMDP SP OP Service Manual 4 4...
Page 48: ...Chapter 5 Software 00 053 101 Revision C SonixMDP SP OP Service Manual 5 4...
Page 124: ...Chapter 9 Network Configuration 00 053 101 Revision C SonixMDP SP OP Service Manual 9 8...
Page 142: ...Chapter 11 Maintenance 00 053 101 Revision C SonixMDP SP OP Service Manual 11 16...
Page 150: ...Chapter 12 Troubleshooting Issues 00 053 101 Revision C SonixMDP SP OP Service Manual 12 8...
Page 166: ...Appendix C Ultrasonix Limited Warranty 00 053 101 Revision C SonixMDP SP OP Service Manual C 2...
Page 168: ...Appendix D Service Drawings 00 053 101 Revision C SonixMDP SP OP Service Manual D 2...