SonixMDP/SP/OP Service Manual 00.053.101, Revision C
Chapter 7: Field Service Components
7-33
7.5.4
Reinstalling the Front Block/Ultrasound Module
Equipment/Tools Required:
• * #2 Phillips screwdriver.
To Reinstall the Front Block/Ultrasound Module:
1.
Attach the new front block/ultrasound module using the #2 Phillips screwdriver to fasten the two (2) hinges (two
(2) screws per hinge).
2.
From the exterior of the modulo, using the #2 Phillips screwdriver, fasten the two (2) screws holding the Side
EMI Strip over the door hinges.
Warning:
Do not perform any internal system maintenance if the UPS breakers are turned on.
Caution:
Always wear a grounding strap when opening and working inside the modulo.
M4x6 Flat Head Screw (x4)
Tool: #2 Phillips screwdriver
Intake Side EMI Strip
UXID: 00.030.340
M4x6 Flat Head Screw (x2)
Tool: #2 Phillips screwdriver
Summary of Contents for SonixMDP
Page 1: ...SonixMDP SP OP Ultrasound System Service Manual...
Page 2: ......
Page 4: ......
Page 10: ...Table of Contents 00 053 101 Revision C SonixMDP SP OP Service Manual vi...
Page 12: ...List of Service Drawings 00 053 101 Revision C SonixMDP SP OP Service Manual ii...
Page 44: ...Chapter 4 Performance Testing 00 053 101 Revision C SonixMDP SP OP Service Manual 4 4...
Page 48: ...Chapter 5 Software 00 053 101 Revision C SonixMDP SP OP Service Manual 5 4...
Page 124: ...Chapter 9 Network Configuration 00 053 101 Revision C SonixMDP SP OP Service Manual 9 8...
Page 142: ...Chapter 11 Maintenance 00 053 101 Revision C SonixMDP SP OP Service Manual 11 16...
Page 150: ...Chapter 12 Troubleshooting Issues 00 053 101 Revision C SonixMDP SP OP Service Manual 12 8...
Page 166: ...Appendix C Ultrasonix Limited Warranty 00 053 101 Revision C SonixMDP SP OP Service Manual C 2...
Page 168: ...Appendix D Service Drawings 00 053 101 Revision C SonixMDP SP OP Service Manual D 2...