Chapter 7: Field Service Components
00.053.206, Revision A
SonixOne Service Manual
7-4
To Import User Data from a Backup:
1.
Ensure the system is plugged in and powered on.
2.
Plug the previously-created backup into one of the six USB ports.
3.
Tap the touch screen
button.
4.
Select
Administrator > System > Import…
.
5.
Check the items to be imported.
6.
Select
OK
to begin the import process or
Cancel
to exit without importing.
Caution:
Ultrasonix does not recommend importing user-defined
Presets
created with a previous software version
as they may not be compatible for use with a more recent software update.
Note:
Use
Select All
to check all items and
Clear All
to clear all checkboxes in one step.
Summary of Contents for SonixOne Q+
Page 1: ...Analogic Ultrasound Service Manual SonixOne Q Ultrasound System...
Page 2: ......
Page 4: ......
Page 8: ...Table of Contents 00 053 206 Revision A SonixOne Service Manual iv...
Page 14: ...Chapter 1 General Information 00 053 206 Revision A SonixOne Service Manual 1 6...
Page 16: ...Chapter 2 System Specifications 00 053 206 Revision A SonixOne Service Manual 2 2...
Page 42: ...Chapter 4 Performance Testing 00 053 206 Revision A SonixOne Service Manual 4 4...
Page 46: ...Chapter 5 Software 00 053 206 Revision A SonixOne Service Manual 5 4...
Page 54: ...Chapter 6 Peripherals and Accessories 00 053 206 Revision A SonixOne Service Manual 6 8...
Page 119: ......