SonixOne Service Manual
00.053.206, Revision A
Chapter 11: Maintenance
11-11
11.3.6 Cleaning/Disinfecting Endocavity Transducers
Endocavity transducers are semi-critical medical devices and must be decontaminated using, at a minimum, High
Level Disinfection.
Clean and disinfect transducers prior to the first exam and following each exam thereafter.
Table 11-4: Endocavity Cleaning/Disinfecting Agents
To Clean/Disinfect a Transducer:
1.
Unplug the transducer.
2.
Wash the transducer head and cable with soap and water to remove any protein buildup; however do not rinse
or immerse the transducer near the strain relief.
3.
Following the manufacturer’s instructions, disinfect the transducer with a recommended disinfecting agent from
4.
Wipe with a clean, dry cloth.
11.3.7 Sterilization
Sterilization of transducers is not possible. Follow the instructions for cleaning and disinfection instead:
• Endocavity transducers:
• Non-invasive transducers:
.and
.
E
NDOCAVITY
T
RANSDUCERS
C
LEANING
/D
ISINFECTING
A
GENTS
Cide
x Act
ivate
d D
ialdeh
yd
e
So
lut
ion 14 day
C
id
ex
P
lus
2
8
d
ay
Cide
x OP
A
Cide
zyme
EC9-5/10
♦
♦
♦
♦
EC9-5/10 GPS
♦
♦
♦
♦
4DEC9-5/10
BPC8-4/10
BPL9-5/55
♦
♦
♦
♦
Caution:
Use only Ultrasonix recommended cleaners/disinfectants (
). They have been tested and
determined safe to use on Ultrasonix transducers.
Failure to follow these instructions may cause damage
and will void transducer warranties.
Note:
Where any transducer (including, but not limited to, an intracavity transducer) is used in a clinical
application of a semi-critical nature (including, but not limited to intraoperative, transrectal, transvaginal,
transesophageal, etc.), ensure the transducer is covered with the appropriate STERILE transducer cover/
sheath which has received regulatory clearance for use. Refer to
Third Party Accessories
in Appendix B
of the relevant
Extended User Manual
for the recommended transducer cover/sheath.
Caution:
Do not allow cleaning solutions to air dry on the transducer.
Note:
Where transducers (non-critical and semi-critical medical devices/equipment) cannot withstand
sterilization, the FDA recognizes the use of a sterile gel and a sterile transducer cover as an acceptable
method of infection control for ultrasound transducers.
Summary of Contents for SonixOne Q+
Page 1: ...Analogic Ultrasound Service Manual SonixOne Q Ultrasound System...
Page 2: ......
Page 4: ......
Page 8: ...Table of Contents 00 053 206 Revision A SonixOne Service Manual iv...
Page 14: ...Chapter 1 General Information 00 053 206 Revision A SonixOne Service Manual 1 6...
Page 16: ...Chapter 2 System Specifications 00 053 206 Revision A SonixOne Service Manual 2 2...
Page 42: ...Chapter 4 Performance Testing 00 053 206 Revision A SonixOne Service Manual 4 4...
Page 46: ...Chapter 5 Software 00 053 206 Revision A SonixOne Service Manual 5 4...
Page 54: ...Chapter 6 Peripherals and Accessories 00 053 206 Revision A SonixOne Service Manual 6 8...
Page 119: ......