Testing the Transducer Board
To Test the Transducer Board:
1.
Unlock and remove the transducer from its current transducer port.
2.
Re-attach and lock the transducer into a different transducer port.
3.
Tap the touch screen
button and select the relevant transducer.
4.
Test the transducer by taking a fresh image. If the line/shadow(s) in the image disappears, then the transducer
board is the problem and the front block/ultrasound module rather than the transducer needs to be serviced.
Figure 7-6: Acceptable Transducer Test Image
Note:
for an example of an acceptable transducer image.
If there is any doubt about the image, contact Ultrasonix Technical Support and forward them a digital copy
of the image in question for verification of the diagnosis.