Chapter 3: System Installation
00.053.129, Revision C
SonixTablet Service Manual
3-8
3.3
INSTALLATION
3.3.1
EMI Filter and Power Pack
3.3.1.1 Voltage Configuration
The diagram below shows the location of the EMI Filter which houses the system's fuses. Refer to
for
details on how to remove the fuse box and replace the fuse(s).
When changing the voltage settings, the EMI filter must be adjusted to the relevant setting and the fuses must be
flipped to match the selected voltage.
Figure 3-2: EMI Filter
Figure 3-3: EMI Filter and System Power Button Location
Caution:
The system voltage setting is configured in the factory. It should not need to be changed in the field.
Main Power Switch
Power Cord (to Wall Outlet) Receptacle
Summary of Contents for SonixTablet
Page 1: ...SonixTablet Ultrasound System Service Manual...
Page 2: ......
Page 4: ......
Page 10: ...List of Service Drawings 00 053 129 Revision C SonixTablet Service Manual ii...
Page 30: ...Chapter 3 System Installation 00 053 129 Revision C SonixTablet Service Manual 3 14...
Page 34: ...Chapter 4 Performance Testing 00 053 129 Revision C SonixTablet Service Manual 4 4...
Page 38: ...Chapter 5 Software 00 053 129 Revision C SonixTablet Service Manual 5 4...
Page 44: ...Chapter 6 Peripherals and Accessories 00 053 129 Revision C SonixTablet Service Manual 6 6...
Page 84: ...Chapter 9 Network Configuration 00 053 129 Revision C SonixTablet Service Manual 9 8...
Page 100: ...Chapter 11 Maintenance 00 053 129 Revision C SonixTablet Service Manual 11 14...
Page 118: ...Appendix C Ultrasonix Limited Warranty 00 053 129 Revision C SonixTablet Service Manual C 2...
Page 120: ...Appendix D Service Drawings 00 053 129 Revision C SonixTablet Service Manual D 2...