SonixTablet Service Manual
00.053.129, Revision C
Chapter 3: System Installation
3-11
3.3.2.2 Changing Fuses
Equipment/Tools Required:
• two (2) 3AG Slo-Blo, 7A/250V fuses
• fine tipped, narrow shafted, flat head screwdriver (e.g,
⅛
" (3mm) jewelers screwdriver).
To Change the Fuses:
1.
Power off and unplug the power cord from both the wall outlet and the system.
2.
Use the screwdriver to open the fuse box lid.
3.
Position the screwdriver at the top (as in
), using it to remove the fuse box.
4.
Replace one or both fuses, as required.
5.
Reinsert the fuse box.
6.
Close the fuse box lid.
Warning:
Do not perform any internal system maintenance if the UPS breakers are turned on.
Caution:
Ensure the correct voltage setting is visible and that it matches the setting on the EMI filter (
Summary of Contents for SonixTablet
Page 1: ...SonixTablet Ultrasound System Service Manual...
Page 2: ......
Page 4: ......
Page 10: ...List of Service Drawings 00 053 129 Revision C SonixTablet Service Manual ii...
Page 30: ...Chapter 3 System Installation 00 053 129 Revision C SonixTablet Service Manual 3 14...
Page 34: ...Chapter 4 Performance Testing 00 053 129 Revision C SonixTablet Service Manual 4 4...
Page 38: ...Chapter 5 Software 00 053 129 Revision C SonixTablet Service Manual 5 4...
Page 44: ...Chapter 6 Peripherals and Accessories 00 053 129 Revision C SonixTablet Service Manual 6 6...
Page 84: ...Chapter 9 Network Configuration 00 053 129 Revision C SonixTablet Service Manual 9 8...
Page 100: ...Chapter 11 Maintenance 00 053 129 Revision C SonixTablet Service Manual 11 14...
Page 118: ...Appendix C Ultrasonix Limited Warranty 00 053 129 Revision C SonixTablet Service Manual C 2...
Page 120: ...Appendix D Service Drawings 00 053 129 Revision C SonixTablet Service Manual D 2...