Using the Ultrawave Cycle Validation Software
If the MC card Validation mode is selected, cycle information is saved to the inserted
MC card and data can be saved in an electronic format. The following pages explain
how to use the MC card and the Ultrawave Cycle Validation Software.
1.
Setting the Validation Mode
This is done in Screen 1 of the
Administrator Instruction
menu (see pg. 4). In order to
enable digital cycle validation set the Validation Mode to MC Card only or Printer & MC
Card.
2.
Inserting the MC Card
In order for data to save to the MC card, the card must be inserted into the port
located on the left hand side of the control panel before your Hygea 2850VM is
switched on.
The digital cycle traceability will save the information of each cycle (as shown on the
ticket on page 23) to the MC card. This information can then be transferred to a PC
and an electronic copy saved, using the Ultrawave Cycle Validation software supplied
on the MC Card.
If the MC card is not inserted, then a message will display on the screen in Operator
mode “MC Card not present”. If MC Card only is se
lected as the Validation mode, and
the MC card not present, then the cycle will not run.
If no MC card is available, then the chosen Validation Mode will have to be amended
in the Administrator mode menu.