background image

 

       

 
 
 

 

 

Operator instructions 

Ensure you 

Hygea 2850VM

 ultrasonic bath is plugged into a 220-

240V fully earthed supply. 
 
Once  plugged  in,  switch  on  your 

Hygea 2850VM

  via  the  power 

on/off at the rear of the unit.  
 
When  the 

Hygea  2850VM

  is  first  switched  on,  the  following  screen  will  appear  for 

approximately 5 seconds. 
 

 

 
While this screen is displayed, the buttons are inoperative. This is normal as the 

Hygea 

2850VM

 is running through its start-up routine and checking all internal systems are 

working correctly. 
The top line of the display indicates the 

Hygea 2850VM

 serial number. This uniquely 

identifies the machine. 
After 5 seconds, the 

Hygea 2850VM

 is ready for use. 

Summary of Contents for Hygea 2850VM

Page 1: ...Hygea 2850VM Ultrasonic Cleaner Operator Instruction Manual...

Page 2: ......

Page 3: ...0 Hygea 2850VM Cycle validation 21 Using the Ultrawave Cycle Validation Software 22 Hygea 2850VM good practice 27 Detergents 28 Cleaning the HYGEA 2850VM 29 Compliance with the Control of Noise at Wor...

Page 4: ...ays runs its automatic de gas cycle see pg 27 for the importance of degassing Always use a basket or instrument cassette and basket carrier to support items to be cleaned Not doing so may damage the b...

Page 5: ...tificate of Test 1 x Declaration of Conformity 1 x Warranty Registration card 1 x Hygea 2850VM Administrator instruction manual 1 x Hygea 2850VM Operator instruction manual How to install Screw the ho...

Page 6: ...accept the cleaning cycle settings Select your operator identification Press the ENTER key to start the cleaning cycle At the end of the cleaning cycle remove the basket holder from the Hygea 2850VM a...

Page 7: ...first switched on the following screen will appear for approximately 5 seconds While this screen is displayed the buttons are inoperative This is normal as the Hygea 2850VM is running through its sta...

Page 8: ...ing screen will display The Hygea 2850VM will automatically start heating and degassing the liquid When the liquid reaches the required temperature the following screen will be displayed and the Hygea...

Page 9: ...ve up and down the list indicating which name is selected Press the Next button to show the next group of four names Up to 20 names can be stored in the operator list When the correct operator is sele...

Page 10: ...page 30 for information about detergents The Hygea 2850VM will then begin the cleaning cycle During the cycle the lid will lock and cannot be opened until the cycle is completed or aborted During the...

Page 11: ...he details of that cycle Once the ticket is printed simply tear off and store as necessary Ultrawave Version V2 00 Model HYGEA 2850VM Serial No 062204 Time 15 30 Date 01 12 18 Dental Clinic Op 01 Nurs...

Page 12: ...ion software supplied on the MC Card It is recommended that the MC Card is removed and the information transferred to the PC at the end of each clinical session twice a day Note in order for the data...

Page 13: ...e screen displays a Low Fluid warning as below the bath will not operate until the unit is filled to the correct level and the detergent dose added If the screen shows a Lid Open warning the cycle wil...

Page 14: ...ng press and hold the enter button at the bottom of the screen until the warning stops If the operator aborts the cleaning cycle the following screen will be displayed If the fluid level drops below t...

Page 15: ...a record on the MC card when the machine is next switched back on The cycle will be recorded an aborted cycle If the cycle fails due to a power failure the following screen is displayed when the power...

Page 16: ...top and the following screen will display If the ultrasonic power drops below a certain point during a cycle the cycle will stop and the following screen will be displayed If the cycle aborts an audib...

Page 17: ...sure that the drain is either directly over a sink or is connected to a tube which is placed into a sink Step 2 Turn the tap until open DRAIN TAP SHOWN OPEN Step 3 Allow the liquid to drain Step 4 Clo...

Page 18: ...paper and fit a new roll Note Hygea 2850VM will only print on thermal paper roll and this must be inserted correctly Step 1 Open the printer cover by pulling the clear cover towards you Step 2 Hold th...

Page 19: ...through the printer by pressing the paper advance button on the right hand side of the printer The Hygea 2850VM is now ready to use When the paper roll is coming to an end a red stripe will appear on...

Page 20: ...nic Energy Meter available from Ultrawave If there is no access to the Energy Meter this test should be replaced by the aluminium foil ablation test 3 Quarterly tests carried out by the operator Estat...

Page 21: ...Therefore after 11 months in operation a reminder will appear on the LCD display warning that the annual validation testing is due The Validation Testing reminder will display as below When this remi...

Page 22: ...ual If the hard copy print out is chosen then at the end of each cycle the integral printer in the Hygea 2850VM will print a ticket showing the details of that cycle Once the ticket is printed simply...

Page 23: ...d the card must be inserted into the port located on the left hand side of the control panel before your Hygea 2850VM is switched on The digital cycle traceability will save the information of each cy...

Page 24: ...IT administrator Also ensure that Microsoft Net 3 5 Framework is installed as well as the associated updates In order to install the Software the MC Card should be inserted into the PC using the USB c...

Page 25: ...e operator to copy the data from the MC Card to the PC If option 1 is chosen and data is already saved to the PC the data will display in the below format If you wish to copy new cycle data to the sof...

Page 26: ...in the Cycle column on the left hand side Then press the View Single Cycle Result button to see a ticket showing the data associated with this cycle If required this can then be printed by pressing th...

Page 27: ...for individual cycle data In order to use the Search function simply type the search field in the applicable search box located in the top right of the screen The following fields can be used to searc...

Page 28: ...te and bake on to the instruments Ultrasonic activity itself will heat up the liquid at a rate of approximately 10 15 C per hour if in continuous use Baskets and Instrument Cassettes Instruments shoul...

Page 29: ...ed to the Hygea 2850VM every time the bath is refilled It is recommended that this occurs twice daily Ultraclean M2 Ultraclean M2 is a general purpose non enzymatic detergent suitable for use on all m...

Page 30: ...ltrasonic activity by vibrating at very high speeds If any contaminants are in contact with the bath they act as an abrasive causing wear on the metal surface In extreme cases the bath will develop ho...

Page 31: ...g is now 80 decibels There is also an exposure limit value of 87 decibels taking account of any reduction in exposure provided by hearing protection above which workers must not be exposed To help you...

Page 32: ...ble acidic caustic or corrosive chemicals or fluids not recommended by Ultrawave The user should familiarise themselves with this instruction booklet before operating the equipment and should apply to...

Page 33: ...s message displays run another cycle to see if the problem persists The screen displays OVER TEMP Drain and refill with fresh cold water The Hygea 2850VM is fitted with a temperature sensor that autom...

Page 34: ...althcare applications should be decontaminated packaged in accordance with MHRA guideline document DB2003 5 Management of Medical Devices prior to Repair Service or Investigation This can be found at...

Page 35: ...and dose sachets Ultraclean M2 detergent A non enzymatic detergent packaged in boxes of 6 x 1 Litre bottles Hygea Ultrasonic Validation Kit A complete Validation kit providing all the material require...

Page 36: ...s Brownes strips and holders for weekly testing of the Hygea 2850VM Service Contracts and Validation Testing Ultrawave can offer full Service and Validation packages for the Hygea 2850VM to ensure the...

Page 37: ..._______________________________ _______________________________________________ _______________________________________________ _______________________________________________ ________________________...

Page 38: ..._____________________________ _______________________________________________ _______________________________________________ _______________________________________________ __________________________...

Page 39: ......

Page 40: ...Service Tel 44 0 845 330 4238 Fax 44 0 845 330 4231 www ultrawave co uk admin ultrawave co uk service ultrawave co uk Ultrawave baths are manufactured in the United Kingdom Ultrawave baths are manufac...

Reviews: