IND Operating Manual
3
C6000507
Safety Instructions
Use of products
The following products are covered by this operating manual:
•
IND30, IND45, IND75, IND80, IND90, IND90L, IND100, IND100L, IND145,
IND145L, IND155L
The products listed above are a series of ultrasonic cleaning baths designed for indoor
use by a professional user.
Electrical:
The mains plug (BS 1363) is fitted with a 13A fuse.
NEVER FIT A FUSE OF
A HIGHER RATING.
The IND machines are rated at up to 3000W
Connect to a 230 VAC fully earthed supply via a BS 1363 3 pin plug socket.
It can be dangerous to operate equipment without an earth connection.
The mains switch is the disconnect device and should remain accessible while
the machine is in use.
Ensure that excess mains cable is stored neatly.
Temperature:
Care should be taken when operating the equipment at higher temperatures as
external surfaces may become hot.
When the temperature is set to 54°C or above, the unit will become hot and care
should be taken when touching the case.
A risk assessment should be conducted
–
see BS EN ISO 13732-1:2008 for full details
The liquid level float switch is the safety device ensuring that the heaters only operate
if there is water in the tank. It should be free to move on its stem. This should be
checked every time the tank is emptied