Ensure to discharge helium gas and remove either side of the self-sealing couplings from
the used adsorber before disposing.
When the refrigerator system is not in operation, helium gas pressure inside the adsorber
is as high as the charge pressure of the compressor. Disposing the adsorber with helium
gas inside may cause accidents. For example, if the adsorber was thrown in a burner
reactor without discharging helium gas by mistake, the inside pressure of the adsorber
rises as the inside temperature rises and may result in explosion. If the adsorber was
pressed with helium gas remaining inside, it may also cause explosion. Refer to
“Disposal Consideration” for more information.
Use appropriate charging adaptor to remove helium gas safely.
☆
We collect used adsorbers. If there is any used absorber, please contact a person in your
local dealership or Ulvac Techno Ltd. to have it picked up or send it directory to our
Service Engineering Division. We only accept CTI adsorbers and our adsorbers.
・
Power supply voltage must be kept within the allowable range(refer to “Section 1 Table1-1
Compressor Unit Standard Specifications” . Applying more than the allowed voltage
may damage the equipment.
・
Install a ground-fault circuit interrupter(rated current: 20A, rated tripping leakage
current:30mA 3
φ
, rated voltage: 200V).
・
Grounding conductor (earth wire) must be connected. (D class grounding (with the
ground resistance of
100Ω)).
・
Connect the refrigerator cable and remote cables (signal lines) to the compressor before
connecting the compressor to the power supply. These wiring connections must not be
done when the compressor is powered on. It could lead to a serious accident such as
electric shock and/or damage to the equipment.
・
Follow recommended sizes for wires and cables.
・
The distance between power line and signal line must be larger than 100mm.
4. Disposal of used adsorber
5. Electrical wiring connections
Summary of Contents for C15T
Page 2: ......
Page 8: ...F 6 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 14: ...S 4 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 16: ...IW 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 34: ...Operation 4 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 38: ...Disconnection and Storage 5 4 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 46: ...Maintenance 6 8 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 50: ...Troubleshooting A 4 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 54: ...Circuit Diagram B 4 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 57: ...Flexible Hose C 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 61: ...CS 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 63: ...Revisions R 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...