Compressor Unit Instruction Manual
ULVAC CRYOGENICS INCORPORATED
3-7
3.5
Connecting Electrical Cables
Do not connect the compressor to power supply until all wiring and connections are
completed.
Disconnect the input power connector at REMOTE/RESPONSE wiring. Failure to observe
this precaution could result in damage to, or destruction of, the equipment.
WARNING
WARNING
1. Make the REMOTE/RESPONSE wiring, if necessary.
・
Do not operate the compressor with excess helium (static charge pressure 1.65MPaG or
above at room temperature 20°C) to improve reliability.
・
The operating pressure tend to be higher with flexible hoses of 3 meters or longer.
The charge pressure for different hose length is as follows;
<Flexible Hose Length> <Charge Pressure>
3 meters (standard)
1.60MPaG
10 meters
1.53MPaG
15 meters
1.50MPaG
20 meters
1.47MPaG
CAUTION
C
har
ge
P
rs
s
u
re
(MP
aG
)
1.65
1.60
1.55
1.50
1.45
3
6 9 12 15 18 21
10
20
Flexible Hose Length (m)
Summary of Contents for C15T
Page 2: ......
Page 8: ...F 6 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 14: ...S 4 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 16: ...IW 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 34: ...Operation 4 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 38: ...Disconnection and Storage 5 4 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 46: ...Maintenance 6 8 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 50: ...Troubleshooting A 4 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 54: ...Circuit Diagram B 4 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 57: ...Flexible Hose C 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 61: ...CS 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Page 63: ...Revisions R 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...