DC Power Supply
【
DC-10-D/DC-20-D/DC-10-DH/DC-20-DH
】
2. Overview
-8-
2-6 Connection of Input/Output Section
2-6-1 Connection of input/output section
[HARTING] connector is employed for the input section.
Although the genuine cable can be purchased from us, if you prepare a cable by yourself, purchase
connector component parts with reference to the items of [Table 2-7: Configuration of the HARTING
connector on the cable side of the input-port], select a suitable wiring material, and complete the
installation.
A special tool is required. Contact the manufacturer for the details of installation conditions etc. When
selecting a wiring material used for the cable, refer to [Table 2-3/Table 2-4: Electrical specifications],
and select the one that would satisfy the rating specification. We recommend that you use our cable
because it is available as an option from our company. In addition, select a suitable cable ground
based on the line shape of the wiring material selected.
・
Input cable: 3-phase + GND, withstand voltage 600V or more, allowable current 80A or more.
Figure 2-
1:
Assignment for the product side of the input connector
Table 2-7: Configuration of the HARTING connector on the cable side of the input-port
Name
Model
Quantity
100A single module female
09 14 001 3131
3
Hinged frame
10B
(
A-C
)
09 14 010 0361
1
10B-gg-R-M40
19 30 010 0448
1
Han TC100 famale contact 16mm
2
09 11 000 6216
3
Han-Mod PE contact extensIOn special cab
09 14 000 9912
1
Table 2-8: Pin assignment of the HARTING connector on the cable side of the input port
Connector / Pin Number
Connection destination
A
R-phase
B
S-phase
C
T-phase
PE(Protective Earth)
GND
Table 2-9: Termination of optional input cable
Connection destination
Our cable color and load terminal
processing
Size of solderless
terminal
R-phase
Black 1 (red insulation cap)
φ
10
S-phase
Black 2 (white insulation cap)
φ
10
T-phase
Black 3 (black insulation cap)
φ
10
GND
G/Y (green insulation cap)
φ
10
A
B
C
PE