35
YK18-0012-AI-101-02
6
.
Turning ON/OFF of P1 filament (electric discharge)
d. By pressing FIL switch on the Main screen, and pressing FIL ON switch on the Filament
ON/OFF screen, a pressure measurement by G-TRAN connected to P1 is started. By pressing
FIL OFF switch, the filament is turned OFF and the pressure measurement is stoppe
7
.
Changeover of P1 filament
If G-TRAN SH2-1 is used, the filament used for the contact signal between CNSI 5-6 pin can
be changed over.
Connector
Contact signal
Description
CNSIL
5-6PIN
Open
Using filament FIL1
Close
Using filament FIL2
Changing the display unit
The display unit can be changed from Pa to Torr or mbar by the following operations.
1
.
Press and hold the ULVAC logo when the power is on.
2
.
Press the following switch
3
.
When the switch display changes, press Return.