background image

 
 

SK00-9663-E8-003-00 

CE standard 

Low Voltage Directive 

EN61010-1:2010(Third Edition),A1:2019 

2014/35/EU 

EMC Directive      EN61326-2-3:2013 

2014/30/EU 

Static electricity test 

         IEC61000-4-2:2008 

Radiation electromagnetic field test 

IEC61000-4-3:2006+A1:2007+A2:2010 

Transient burst test 

IEC61000-4-4:2004+A1:2010 

Lightening surge test 

IEC61000-4-5:2005 

Conduction test  IEC61000-4-6:2008 

Commercial magnetic field test 

IEC61000-4-8:2009 

Radiation field intensity measurement 

CISPR11:2009+A1:2010:Group 1 Class A 

RoHS        EN IEC63000:2018 

2011/65/EU 

UKCA 

standard

 

Conforms to British legislation equivalent to EU 

Directives for CE standards listed above. 

I/O connector 

D-sub15pin connector (male), M2.6mm screw 

USB Type-C 

USB Type-C can only be used for communication using UL-

MOBI. 

Weight 

SW100 only: Approx.105g 

Dimensions 

SW100 only: Approx.48mm×30mm×74mm 

See Section 6.1 

Sensor 

See 

Section 

6.2 

Model 

SWP-16 

SWP-25 

SWP-R1/8 

SWP-P18 

SWP-P15 

SWP-CF16 

SWP-1S 

Inner 

volume 

7.3cm

3

  7.8cm

3

  7.6cm

3

  13.9cm

3

  10.0cm

3

  11.3cm

3

  8.2cm

3

 

Weight 

45g 

48g 

44g 

43g 

30g 

60g 

95g 

accuracy

1

: The accuracy after atmospheric pressure and zero  point adjustment. 

Please carry out atmospheric pressure and zero point adjustment before 

use especially. Moreover, since calibration gas is N

2

, when other gas 

is measured, be careful of an error of measurement. 

Summary of Contents for G-Tran Series

Page 1: ...eries Pirani Vacuum Gauge Sensor Unit Analog Output Type Model SW100 A Serial Communication Type Model SW100 R Specification ULVAC Inc Components Division 253 8543 2500 Hagisono Chigasaki Kanagawa Jap...

Page 2: ...For details refer to the SW100 instruction manual USB Type C can be connected to a Windows Android device with UL MOBI installed using the USB cable and can be monitored and set by UL MOBI UL MOBI fo...

Page 3: ...input signal Zero point atmospheric pressure adjustment adjustment reset Operates with open collector input negative logic Control output signal Sensor error signal setpoint1 2 Open collector output...

Page 4: ...eating temperature Sensor only 150 without SW100 When heating deviations from specifications such as accuracy is observed The operating temperature range of SW100 is 10 to 40 Mounting position Free IP...

Page 5: ...nforms to British legislation equivalent to EU Directives for CE standards listed above I O connector D sub15pin connector male M2 6mm screw USB Type C USB Type C can only be used for communication us...

Page 6: ...unwired Calibration certificate Calibration certificate JCSS calibration certificate Inspection certificate Traceability certificate Display unit Model ISG1 DC24V power supply Display cable Between SW...

Page 7: ...00 for both new installation and replacement and set and connect When used as a compatible unit the performance of SW100 A may be limited so please read the instruction manual UL MOBI can be downloade...

Page 8: ...MAX 70mW 5 Adjustment input Input a signal when adjusting zero point or atmospheric pressure Operates when short with GND 8 Measure signal output It outputs a measured value DC 0V to 10V 9 Power suppl...

Page 9: ...70mW 4 RS232C RxD RxD of RS 232C 5 Terminating resistor for RS485 Terminating resistor for RS 485 Connect with pin13 6 RS232C TxD TxD of RS 232C 8 Measured value output It outputs a measured value DC...

Page 10: ...ted as OUTPUT see Section 1 4 of this manual follow the respective conversion formulas instead of the pressure conversion formulas in this section For details refer to the SW100 instruction manual Mea...

Page 11: ...rcuit diagram Control input signal Analog output type SW100 A only Fig 3 2 SW100 A input signal internal circuit diagram 5V 24V A 1 2 A 1 2 A 1 2 2 ERROR signal ON OFF 3 SETPOINT signal 1 ON OFF 15 GN...

Page 12: ...connection 1 Maximum connections 32 including host 9600 19200 38400 bps 9600 19200 38400 bps Settings RS 232C connection diagram An example of RS 232C connection is shown below The host side is a D s...

Page 13: ...0 AD1 CMD D0 Dn SH SL CHKH CHKL CR Colon AD0 Device address higher 4bit 0 9 AD1 Device address lower 4bit 0 9 CMD Commands upper lower case sensitive D0 Data 4bit 0 9 A F Dn Data 4bit 0 9 A F SH Highe...

Page 14: ...te Setpoint1 value SR Reading status 2W Write Setpoint2 value SW Writing status Command Excerpt Reading measured value and status Command AD0 AD1 D CHKH CHKL CR Return format from this unit to PC AD0...

Page 15: ...is damaged due to problem during transportation at the time of delivery The warranty scope specified in the latest INCOTERMS shall be applied 3 A product that does not meet the basic specifications o...

Page 16: ...party s claim that we infringe a patent 8 Sensor unit in use Lifespan due to the use measurement error due to contamination 9 Sensor cable in use Cable disconnection or poor contact due to improper in...

Page 17: ...SK00 9663 E8 003 00 Diagrams...

Page 18: ...3 00 SWP Sensor for SW100 dimensions 46 40 30 SWP 16 NW16 66 60 47 41 40 77 83 18 SWP 25 NW25 SWP R1 8 R1 8 SWP P18 18tube 45 51 50 4 65 15 59 65 59 33 8 SWP P15 15tube SWP CF16 CF16 ICF034 SWP 1S ASM...