3.6 Water Piping
This pump, cooling water is required(Air-cooled type, cooling water is not required)
Connect piping to the Cooling water inlet / outlet using care not to mistake the port.
(Refer to Fig.8 ~ 12.)
MODEL
Cooling
method
Cooling water volume
Cooling water
temperature
Connections
PMB100D
Air cooling
PMB300D
Water cooling
2
L
/min
≦
setting
5
℃
- 30
℃
Rc3/8
PMB600D
Water cooling
2
L
/min
≦
setting
PMB1200D
Water cooling
3
L
/min
≦
setting
PMB2400D
Water cooling
3
L
/min
≦
setting
Rc1/2
CAUTION
Be sure to flow the cooling water with the volume indicated on the
Table. 2 or more. The pump temperature would rise up if the water
volume becomes less than specified particularly on high inlet
pressure operation and cause the pump failure. You are
recommended to put a Flow meter for the Coolant and cause the
interlock so that the Pump stops if the cooling water flows less than
specified volume.
When the operation stopped in winter, the water piping and the
pump have a risk of breaking by freeze-up of the water inside. Open
the cooling water outlet during operation stop and storage to
discharge the water inside by blowing in the pneumatic air through
the cooling water inlet.
CAUTION
If you use several pumps, be sure to connect the cooling water pipes
parallel. Cooling capacity might come down if connected them serial
and cause the failure.
You should put a filter at the front stage if you are obliged to use the
water containing much impurity such like water stain, iron and the like.
An enough cooling water flow might not be able to be secured when
there is a vertical interval in piping. In such a case, cause a measure to
ensure the flow volume, such as to change the piping layout, put a larger
pipe or raise the supply pressure within the specification range.
Summary of Contents for PMB100D
Page 2: ......
Page 3: ...Declaration of Conformity...
Page 39: ...YK11 0016 DI 002 08 12 2 4 Dimensional drawing Fig 11 Dimensional drawing PMB100D...
Page 40: ...YK11 0016 DI 002 08 13 Fig 12 Dimensional drawing PMB300D...
Page 41: ...YK11 0016 DI 002 08 14 Fig 13 Dimensional drawing PMB600D...
Page 42: ...YK11 0016 DI 002 08 15 Fig 14 Dimensional drawing PMB1200D...
Page 43: ...YK11 0016 DI 002 08 16 Fig 15 Dimensional drawing PMB2400D...
Page 62: ...YK11 0016 DI 002 08 35...
Page 65: ...YK11 0016 DI 002 08 38...
Page 72: ...YK11 0016 DI 002 08 45...
Page 84: ...YK11 0016 DI 002 08 57 Fig 28 PMB300D Optional attachment figure...
Page 85: ...YK11 0016 DI 002 08 58 Fig 29 PMB600D Optional attachment figure...
Page 86: ...YK11 0016 DI 002 08 59 Fig 30 PMB1200D Optional attachment figure...
Page 89: ...YK11 0016 DI 002 08 62 Fig 33 Dimensional drawing PMB2400D Horizontal exhaust model...
Page 99: ...YK11 0016 DI 002 08 72...
Page 105: ...YK11 0016 DI 002 08 78...
Page 119: ......