14

 

ULVAC CRYOGENICS INCORPORATED

 

3.3.2 Remote Connector 

The remote connector can be used to remotely operate of the compressor unit. 

Table 3-5 shows the detail of remote operation and monitor sequence. 

 

Table  3-5 

  Remote/Monitor Sequence

 

No. 

Item 

Status 

Pin No. 

Contact Capacity 

Remote 

Operation 

Stop 

Release 

A, B 

Rated Volume: AC200V

 

 

above 0.1A 

Minimum load: AC 200

 

0.01

 

Run 

Connect 

 

Table  3-6  

Applied connector for remote

 

Name 

Specifications 

Plug connectors 

Straight type

 

Made by Japan Aviation Electronics Industry, Ltd. N/MS3106B-14S-7P 

Plug connectors 

Angle type

 

Made by Japan Aviation Electronics Industry, Ltd. N/MS3108B-14S-7P 

Cable clamps 

Made by Japan Aviation Electronics Industry, Ltd. N/MS3057-6A 

 

Diagram 1.5 Remote connector pin configuration 

 

 

 

Figure 3-7  Remote Connector Pins 

 

3.3.3   Signal Connector 

The signal connector can be used to monitor conditions. 

Table 1.7 shows the monitor sequence of the compressor unit. 

Table  3-7 

Monitor sequence 

No. 

Items 

Condition 

Pin code 

Volume of contact points 

Operation monitor 

No-voltage 

contact 

point 

Operated  Close 

Stopped  

 

Open 

1 , 2 

DC 24V  0.1A 

Warning monitor 

No-voltage 

contact 

point 

Normal   Open 

Warning  Close 

4 , 5 

DC 24V  0.1A 

 

Table  3-8 

Applicable connector for remote

 

Name 

Specifications 

Plug connector 

DE9S-NR    Made by Japan Aviation Electronics Industry, Ltd.   

Connector Shell 

DE-C8-J9-B1-1R    Made by Japan Aviation Electronics Industry, Ltd.   

   

  Signal connector pin configuration 

 

 

 

Figure 3-8  Remote Connector Pins  

 

This is the socket side insert, as seen from the pins at 

front of connector. 

This is the socket side insert, viewed  from 

the pins at front of connector. 

Summary of Contents for SA115-C

Page 1: ...nsors the cryocooler falls under row 6 A 2 d 2 of the control list established by The Wassenaar Arrangement which is equal to row 10 2 of appended table 1 of Japan s Export Trade Control Order Custome...

Page 2: ......

Page 3: ...manual close at hand so that you can use for reference during operation If you purchased our other products and or optional devices with this product read relevant instruction manuals carefully Genera...

Page 4: ...unsafe practices which could result in severe injury or loss of life A caution describes safety hazards or unsafe practices which could result in personal injury or equipment damage Toxic gas or chem...

Page 5: ...Instructions for the Electrical System 11 4 Installation 17 4 1 Conditions for Installation 17 4 2 Installation 18 5 Operation 22 5 1 Start Operation 22 5 2 Stop Operation 23 5 3 Checking Normal Oper...

Page 6: ...IV ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...

Page 7: ...r repair or disposing Do not use this device in the corrosive gas atmosphere e g chloride It may result in injury or damaging devices Internal parts such as compressor motor or exhaust piping are in e...

Page 8: ...charging adaptor for refrigerator maintenance 2 Before proceeding with disposal work remove the couplings of the adsorber to visually confirm that it has been depressurized Disposal Considerations The...

Page 9: ...exterior b Oil leaks c Helium gas charge pressure The pressure gauge indicates the helium charge pressure of 1 9MPa at 20 C helium charge pressure may vary depending on the system configurations If t...

Page 10: ...gas With purity of 99 999 or above Refrigerant filling pressure MPa 1 90 0 05 20 C Dimensions Height mm 484 Width mm 494 Depth mm 494 Weight kg 59 Safety Valve Setting MPa 2 84 3 24 Connec tion Refrig...

Page 11: ...ally continue 2 HOUR METER HM The time counter indicates the total hours of operation of the compressor unit 3 SUPPLY PRESSURE Pressure meter When not in operation indicates the helium gas charge pres...

Page 12: ...6 ULVAC CRYOGENICS INCORPORATED Figure 3 1 SA115 C Control and Connection Components NAME PLATE LABEL 484 460 5 5 58 426 494 85 259 70 100 391 381 167 167 66 8 5 160 59 59 376 60 494...

Page 13: ...lium gas and oil in compressor unit SA115 C Figure 3 2 SA115 C Helium gas flow Helium Flow Oil Flow Oil Separator Pressure Regulator Refrigerant Low Pressure Switch Compressor Pressure Equalizing Sole...

Page 14: ...ating oil 12 Low pressure regulator By pass valve to maintain the pressure of the low pressure helium gas above a certain level 13 High pressure regulator By pass valve to maintain the pressure of the...

Page 15: ...ULVAC CRYOGENICS INCORPORATED 9 Figure 3 3 SA115 C Internal Components 1 18 12 22 23 3 2 4 11 21 13 6 7 5 14 10 15...

Page 16: ...10 ULVAC CRYOGENICS INCORPORATED Figure 3 4 SA115 C Internal Components 9 8 17 19 20 24 16...

Page 17: ...Terminal Board TB1 KTU15 J 600V 15A 4P 1 8 Thermostat 26G CS 74L120 120 C 2 9 Low Pressure Switch 63PL ACB 1AU3 OFF 0 098MPa 1 10 Reverse Phase Prevention Relay 47C QE 20N AC200V 1 11 Magnet Switch T...

Page 18: ...12 ULVAC CRYOGENICS INCORPORATED Figure 3 5 SA115 C Internal Components 7...

Page 19: ...ULVAC CRYOGENICS INCORPORATED 13 Figure 3 6 SA115 C Internal Components 8 9 15 16 11 12 14 13 10 17...

Page 20: ...mote connector pin configuration Figure 3 7 Remote Connector Pins 3 3 3 Signal Connector The signal connector can be used to monitor conditions Table 1 7 shows the monitor sequence of the compressor u...

Page 21: ...ray A L1 B L2 C L3 D GND Table 3 10 Coldhead Connectors Item Specifications Plug Connector Straight Made by Japan Aviation Electronics Industry Ltd JL04V 6A 18 10PE EB R Plug Connector Angle type Made...

Page 22: ...sure drops due to some reasons such as the amount of helium gas inclusion is less than the specified amount Solenoid Valve Stabilize the helium pressure between the supply and return flexible hoses wh...

Page 23: ...upply line are needed Power Source Make sure that the correct AC power source is supplied to the compressor For more information see Table 3 1 Compressor Unit SA115 C Specifications Room Temperature C...

Page 24: ...plate 4 2 2 Install the Compressor Unit Check that the following installation conditions are met 1 Install the compressor unit on a smooth floor that is visually level 2 Be sure that all four casters...

Page 25: ...2 and IEC60947 3 Install a circuit breaker in a place close to the compressor unit within 3m and can be easily accessed Do not install in a way that hinders easy operation Indicate that the circuit b...

Page 26: ...m2 up to 30 m 3 5 mm2 Electric wiring Room temperature 38 C or less Metal tube PVC tube etc Voltage drop standard 2 Minimum thickness of power cable 2 mm2 21 m Distance up to 10 m 2 mm2 up to 20 m 2 m...

Page 27: ...the female self sealing coupling of the high pressure piping outgoing piping to the refrigerator to the male self sealing coupling of the SUPPLY and screw the union nuts of the female self sealing cou...

Page 28: ...ables are correctly connected To start the cryocooler system 1 Confirm that the pressure of the compressor unit is within the rated values 2 Turn the OPERATION switch to ON on the front panel of the c...

Page 29: ...The ultimate temperature of the coldhead depends on the heat load or vacuum level of the vacuum chamber If the temperature is high and stable the total heat load for the coldhead may be too large Chec...

Page 30: ...iability of cryocoolers It is requested to clean the heat exchanger at least once in a circumstance such as computer rooms The timing may vary depending on the ambient conditions of a compressor Clean...

Page 31: ...d connect helium charging valve to the port 4 Connect the charging hose to the pressure regulator of the helium bottle 5 Connect the charging hose to the charging valve while releasing small amount of...

Page 32: ...fications Component No Notes F1 F2 F3 Glass tube fuse 250VAC 3A MT4 3A For control circuits NOTES 1 When the helium gas pressure meter reads 0 MPa air may already present Contact our service departmen...

Page 33: ...e 1 Loosen the cap of the fuse holder with a flathead screwdriver and remove the used fuse 2 Insert a new fuse into the fuse holder and fasten the cap as before Disconnect all the power supply to the...

Page 34: ...E LOCAL switch to L or perform the correct remote operations Either the fuse is blown or disconnected Check if the inner fuse is OK If defective replace the fuse with a new one If the problem continue...

Page 35: ...ar if blocked Secure the space to intake and exhaust air around the compressor unit In case the thermostat is activated due to the abnormal temperature it may take time to return to the normal conditi...

Page 36: ...ntact us Compressor unit stops during operation Helium gas temperature is not in normal range A Abnormal temperature The flow rate of the cooling air is low B Lubricating oil of the compressor is decr...

Page 37: ...1 Connectors attached 3 Remote Connector 1 For remote operation 4 Single Head Wrench 1 set For connecting helium gas lines 5 Instruction Manual 1 This book 8 2 Optional Parts Table 8 2 Optional Parts...

Page 38: ...OTE VT WT PK EXTERNAL OPERATION VT BN 9 OR RD RD F1 3 B C DRIVE MOTOR FAN MOTOR MOTOR COMPRESSOR T S R 51 CM PK BN C PK BN CM 52 BL WT RD 47 C 3 4 6 1 2 SW2 L O C A L R E M O T E BN BN VT AL 4X 1 10 2...

Page 39: ...cases the cost shall be your own expense even though the product is within the warranty period i Breakdowns due to improper storage or handling careless accident software or hardware design by the cus...

Page 40: ...ENICS products and compensation for damages to products other than ULVAC CRYOGENICS products and other services are not covered under warranty 3 Repair period after production is discontinued ULVAC CR...

Page 41: ...Chigasaki Kanagawa 253 0085 Japan Sales Tel 81 467 85 8884 Service Engineering Division Tel 81 467 85 9366 Fax 81 467 83 4838 ULVAC CRYOGENICS KOREA INC www ulvac cryo co kr 107 Hyeongoksandan ro Che...

Page 42: ...CS 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...

Page 43: ...CORPORATED R 1 Revision History Date Revision No Contents 2017 07 27 2021 07 08 2017 07 2021JU01 First edition Table 3 4 3 8 3 10 Parts model name has been changed Table 3 1 has been modified Figure9...

Page 44: ...R 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...

Reviews: