7.3.
Scheduled adjustment
There are various environments such as when the ambient temperature fluctuates greatly
or when a lot of dirt such as oil adheres, so it is up to the customer to decide how
often the adjustment should be made.
Tests at ULVAC showed that the same accuracy as that at the start of operation can be
maintained by conducting adjustment about once a month under the following conditions.
That is, this unit is installed directly above the GVD-135A rotary pump, repeating
venting to atmospheric pressure and evacuation at a cycle of 20seconds, and making
adjustment indoors (at room temperature).