33
5.3
Interlock for Vacuum System
5.3
Interlock for Vacuum System
(1) When using, as a backing vacuum pump, a vacuum pump with no check mechanism
(backstream flow prevention) such as dry vacuum pump, etc., install a forevacuum valve
between the turbo molecular pump and the backing vacuum pump to prevent rapid inverse
flow of exhausted gas. And close the forevacuum valve before the backing vacuum pump
stops. (Refer to Fig. 5-5)
(2) If "ALARM" signal is emitted or ALARM lamp lights, shut down the backing vacuum pump
or close the forevacuum valve immediately. Furthermore, when main valve is installed
between the turbo molecular pump and the vacuum chamber, close this valve, too.
5.4
Notes on transportation
This product is precision equipment. Do not give any strong impact or continuous vibration in
transportation, otherwise the product could be damaged. In transportation, please use a means
of transportation which have vibration-proof function (an air suspension truck, for example).
Especially when passing by rough road, we recommend that the product is transported keeping
the packing condition when it ships from ULVAC.
When the product is put on the high temperature / humidity environment for a long time, it
causes the breakdown of the product due to corrosion of mechanical parts or performance loss
of electrical parts. Please transport or store the product under an appropriate environment
Summary of Contents for UTM300B
Page 8: ...vii Location of label...
Page 11: ...x This page is intentionally left blank...
Page 17: ...xvi Table of contents This page is intentionally left blank...
Page 51: ...34 SECTION 5 INSTALLATION This page is intentionally left blank...
Page 70: ...GAS PURGE 7...
Page 77: ...60 SECTION 8 TURBO MOLECULAR PUMP RECONDITION This page is intentionally left blank...
Page 111: ...A 26 Appendix A COMMUNICATIONS This page is intentionally left blank...