8.2
Turbo Molecular Pump Decontamination
8.2
Turbo Molecular Pump Decontamination
All expenses incurred with the decontamination of the turbo molecular pump are the
responsibility of the customer.
8.3
Bearing Replacement
The ball bearing (Fig. 3-1 (6)) and the touch-down bearing are the component of this pump that
is subjected to friction and wear.
The replacement of (the ball bearing and) the touch-down bearing is done only by ULVAC
or an approved service company.
Recommended bearing exchange period depends on mounting direction and gas load.
Table 8-1
Recommended bearing exchange period
Gas load
Mounting direction
Bearing exchange
period
No
Vertical (Note 1)
4 years
Directions other than
vertical direction
3.5 years
Yes
Any direction
2 years
(Note 1) Vertical directions is within
±
5
°
.
(Note 2) It is recommended period and is not an exchange period to be guarantee.
8.4
Check of the rotor blades
Rotor blades of turbo molecular pump are high-speed rotor made of aluminum alloy. It has the
possibility that material strength deteriorates, specially when corrosive gas is evacuated. Regular
check (Customer is liable for the cost.) by ULVAC or a ULVAC approved/authorized service
center is suggested. (Every 1 year is recommended.) ULVAC and/or the service center
investigate rotor blades in every overhaul task and check and suggest a rotor replacement to
customers if any cracks are found out.
Summary of Contents for UTM300B
Page 8: ...vii Location of label...
Page 11: ...x This page is intentionally left blank...
Page 17: ...xvi Table of contents This page is intentionally left blank...
Page 51: ...34 SECTION 5 INSTALLATION This page is intentionally left blank...
Page 70: ...GAS PURGE 7...
Page 77: ...60 SECTION 8 TURBO MOLECULAR PUMP RECONDITION This page is intentionally left blank...
Page 111: ...A 26 Appendix A COMMUNICATIONS This page is intentionally left blank...