  5. Troubleshooting 

Situation 

Cause 

Countermeasure 

 

Room temperature is too high.

 

 

Adjust the room temperature to 
25°C or less. 

 

Used  for  a  short  time  directly 
after 

installation 

or 

after 

prolonged disuse.

 

 

Let  the  device  run  for  24  to  48 
hours and then recheck. 

 

Examine  the  area  around  the 
affected 

part 

before 

the 

maximum pressure is affected. 
Ex: Hermetic, etc.

 

 

There is a leak. 

 

 

Clean  the  glass  bell  jar’s 
O-ring.   

 

Oil  from  the  oil  diffusion  pump 
has degraded. 

 

 

If  air  has  accidentally  been 
allowed  to  enter  or  if  there  is 
some  other  obvious  reason  for 
the  oil  to  degrade,  replace  the 
oil. 

 

Coolant water is too hot. 
Insufficient cooling water. 

 

Set  the  water  temperature  to 

20°C. 

Maintain supply of 1.5 L/min. 

 

Using  a  sample  which  emits  a 
large amount of gas. 
Ex: Plastic base material, etc.

 

 

Change the sample material.   
Extend the evacuation time. 

 

Oil-sealed rotary vacuum pump 

is defective.

 

 

Refer  to  the  oil-sealed  rotary 
vacuum pump user manual. 

 

  Max. pressure low 
    or 
  Max. pressure unsteady 
    or 
  Evacuation power reduced

   

Measurement 

device 

is 

defective.

 

  Replace the device. 

 

Oil  diffusion  pump  does  not 
warm up 

 

  Heater broken.   

 

  Replace the heater. 

Evaporation 

source 

short 

circuit.

 

Replace evaporation source. 

 

Current  does  not  flow  to  the 
electrode  even  when  the 
electrical  evaporator  control 
is used 

 

Electrode  short  circuit  inside 
the feed through collar. 

 

Repair the location of the short 
circuit 

and 

restore 

the 

insulation. 

 

Oil-sealed 

rotary 

vacuum 

pump  thermal  overload  relay 
of  electro-magnetic  contactor 
moves.

 

 

Overloading.   

 

 

Refer  to  the  oil-sealed  rotary 
vacuum pump user manual. 
Manual reset .

 

 

Rotate  the  handle  clockwise 
until  is  grows  heavy,  then 
tighten more. 

 

Clean the main valve seat seal. 

 

The  Pirani  vacuum  gauge 

pressure is reduced when the 

bell  jar  is  vented  to  the 

atmosphere

 

 

Main valve shutoff defective.   

 

 

Overhaul 

the 

main 

valve. 

(handled at our location) 

 

Summary of Contents for VPC-1100

Page 1: ...ease read this manual thoroughly to ensure safe and effective operation of this equipment Keep this manual in a safe place Please note that due to performance upgrade the equipment described in this m...

Page 2: ...ent specifications 3 5 Standard accessories 5 6 How to use the switches handle and operation lever 6 7 Layout of the switches handle and operation lever 7 Flow chart 8 Wiring diagram 8 3 Unpacking and...

Page 3: ...repairs 20 1 Dangers and safety guidelines during maintenance and repairs20 2 Maintenance and repairs by the customer 21 3 Removing maintaining and installing equipment 21 3 1 Oil sealed rotary vacuum...

Page 4: ...t permission of Ulvac Kiko Inc Read this section before using the VPC 1100 Follow the instructions below to safely use the device and prevent personal injuries from occurring Please comply with them a...

Page 5: ...o the nearest terminal If the unit is not properly grounded you may get an electrical shock due to a malfunction or short circuit Check cable capacity If possible do not use an extension cord If using...

Page 6: ...operation and for 30 minutes after being switched off the oil diffusion pump and oil sealed rotary vacuum pump are extremely hot Do not touch directly with your hands May cause burns Use an oil mist t...

Page 7: ...perly or may affect its durability Do not attempt alone This device should be moved by two or more people May cause back pains or other injuries Do not attempt alone Attaching or removing the hoist sh...

Page 8: ...ber gloves or other protection over exposed areas of skin A droplet coming into contact with your skin can be extremely painful Venting After stopping the oil sealed rotary vacuum pump open the RP ven...

Page 9: ...ed exhaust resistance can build up and cause oil leakage from the shaft seal or damage to the oil level gauge Follow all regulations Waste oil used in the pump must be disposed of in accordance with a...

Page 10: ...caused by modifications or repairs carried out by a party other than the manufacturer or by a service company not approved by the manufacturer e Malfunctions caused by noise electric disturbance f Ma...

Page 11: ...upply cut off Prevents the oil diffusion pump from overheating Pressure switch detects cutoff Oil diffusion pump heater interlock Alarm warns if water supply cut off Auto reset when supply restored Si...

Page 12: ...re the standard specifications Please refer to the VPC 1100 Special specifications for details of special configurations 4 Individual equipment specifications Equipment Name Type Specifications Number...

Page 13: ...irani vacuum gauge Power consumption 0 55 kW Oil SMR 100 1 1 L 2 5 Main valve Type SBVM 6AX Butterfly valve 1 6 Liquid nitrogen trap Bore diameter 150 A Injection amount Approx 3 5 L Including vapor 1...

Page 14: ...cifications manual for details 5 Standard accessories 1 Power cable 200 V three phase crimp style terminal 4m 100 V single phase with plug 4m 1 1 2 Hermetic seal port Sealing plate Socket 3 3 3 Gauge...

Page 15: ...n Move counter clockwise Close Move clockwise Fully open Rotate the handle counter clockwise until it grows heavy then turn back a quarter turn Fully close Rotate the handle clockwise until it grows h...

Page 16: ...7 Layout of the switches handle and operation lever...

Page 17: ...Flow chart Wiring diagram GLD 202A...

Page 18: ...in body of the device including casters hoist assembled and box of accessories will be delivered The hoist will be attached on site 3 Installation Warning Check environment This device should be insta...

Page 19: ...leads are green cables 100 V plugs use a grounded plug If you are using a plug adapter connect the ground lead to the nearest terminal If the device is not properly grounded you may get an electrical...

Page 20: ...x Outer diameter 15 mm Caution Secure pipe The coolant pipe should be fixed in place with a hose band Prevents accidents caused by cooling water leakage Note The Japan Refrigeration and Air Condition...

Page 21: ...and insert it into the hoist mounting plate 3 Secure hoist Move the hoist s upper arm so it faces the glass bell jar Secure the hoist using the flat washers spring washers and hexagonal bolts in that...

Page 22: ...in place 7 Removal Perform steps 5 to 2 in reverse order 8 Attaching the liquid nitrogen intake pipe Attachment 1 Preparation Remove the liquid nitrogen intake pipe and funnel cap bracket from the bo...

Page 23: ...y valve lever is set to CLOSE Make sure the RP VENT is fully open 2 Check Switch on RP1 The rear pump will start up If the pump is rotating in the opposite direction you will soon hear a rattling soun...

Page 24: ...ves When handling liquid nitrogen wear rubber gloves or other protection over exposed areas of skin A droplet coming into contact with your skin can be extremely painful Venting After stopping the oil...

Page 25: ...rt 3 2 Operation Exhaust system warm up 1 Your breaker ON Receiving light on main breaker Should light up 2 Main breaker ON 3 Cooling water ON 4 RP1 RP2 ON 5 Three way valve FORE 6 Pirani vacuum gauge...

Page 26: ...F Steps 8 and 9 should be performed when possible Usually the procedure is complete at the end of step 6 4 Using the evaporator 4 1 Opening the bell jar 1 If using an ionization vacuum gauge filament...

Page 27: ...Confirmation 5 Turn on the evaporator power switch 6 The current is gradually thrown to the evaporation source pushing the UP key 7 Once the evaporation element is red hot and the vapor source has sta...

Page 28: ...on 6 4 Water supply temporarily cut off All machinery will automatically return to its state prior to the cutoff 6 5 Water supply cut off for a prolonged period Stop the device 1 If using an ionizatio...

Page 29: ...o fall Follow all regulations Waste oil used in the pump must be disposed of in accordance with all relevant laws and regulations Please handle waste oil in accordance with the laws and regulations If...

Page 30: ...towards you disconnect the motor wiring connectors 3 places by pulling them to the left or right Disconnect the ground lead from the mounting frame Remove the vibration proof rubber nuts 4 places Get...

Page 31: ...racket Remove the oil diffusion pump by pulling it towards you 3 Replace oil See the separate ULK 06A oil diffusion pump user manual 4 Attachment Place the oil diffusion pump underneath the liquid nit...

Page 32: ...ase As thinly as possible so that a film forms Clean the inside of the O ring groove Fit the O ring inside the groove 3 5 Glass bell jar 1 Necessary equipment dust mask gloves 2 Removal Loosen the dro...

Page 33: ...ssary Oil sealed rotary vacuum pump Replace the oil RP VENT closed Three way valve set to CLOSE start RP 1 and RP 2 at the same time If the Pirani vacuum gauge pressure is 5 Pa or higher replace the o...

Page 34: ...uation time Oil sealed rotary vacuum pump is defective Refer to the oil sealed rotary vacuum pump user manual Max pressure low or Max pressure unsteady or Evacuation power reduced Measurement device i...

Page 35: ...eing used accidentally Precautions when using only one pump 1 Do not operate the device while the vacuum hose of the malfunctioning oil sealed rotary vacuum pump is connected Oil or air may back up an...

Page 36: ...ng for suction flange V 175 NBR 1 OK O ring for leakage valve N 6 NBR 1 OK Upper portion conversion flange O ring for leakage valve N 8 NBR 1 OK O ring for suction flange V 175 Attached component FPM...

Page 37: ...10A 4 Bell jar cover Rubber lip KV 1297 01 143 EPDM 0 31m OK The numbers listed above are for the standard specification If optional parts are attached these numbers will differ 6 Disposal 1 Precautio...

Page 38: ...events contamination On site Non stick board Prevents evaporation from sticking to the glass bell jar On site UFC070 adapter To fit an 070 Conflat flange On site Measurement rack Built in ionization v...

Page 39: ...coil For heating the inside of the bell jar Factory Film formation controller For measurement and control of film thickness and film formation speed On site Film thickness sensor Sensor for film cont...

Reviews: