SK00-4390-DI-002-11
44
6.5 Trouble Check List
Table. 7 Trouble check list
Trouble
Causes
Processing method
Reference
The Pump does not
run.
(1) Motor connection is wrong.
(2) Safety circuit such as the
Electromagnetic breaker is
not correctly set.
(3) Oil viscosity got higher.
(4) Foreign substance entered
in the Pump caused
burning the Rotor or the
like.
(5) Reactive agent
accumulated inside the
Pump while the Pump was
stopped after having
exhausted the reactive gas.
(1) Check the connection.
(2) Make the Safety circuit
conform to the Motor
specification.
(3) Replace the oil.
(4) Conduct the overhaul
(replacement of the
Cylinder, Rotor, Cover
and so on.)
(5) Conduct the overhaul
(cleaning inside the
Pump, removal of
reactive agent and so on.)
3.6
3.6
3.2 , 4.5
6.2.2
Trouble
Causes
Processing method
Reference
Abnormal noise
sounds.
(1) Motor rotation direction
is reverse.
(2) Slow leak valve is
clogged.
(3) Rattling sounds on
starting or stopping the
machine.
(1) Do the connection again
to correct the rotation
direction.
(2) Remove the Side panel
and wash and adjust the
Slow leak valve.
(3)There is no particular
problem as it’s a
phenomenon caused by
vanes that temporarily
make irregular motions.
3.6
6.2.10
Summary of Contents for VS1501
Page 2: ......
Page 13: ...SK00 4390 DI 002 11 xi Fig 1 Warning Label display position...
Page 27: ...SK00 4390 DI 002 11 7 Fig 2 VS1501 Dimensional drawing...
Page 28: ...SK00 4390 DI 002 11 8 Fig 3 VS2401 Dimensional drawing...
Page 36: ...SK00 4390 DI 002 11 16 Fig 5 Suction Exhaust piping connection diagram...