OPERATING INSTRUCTIONS
USER MANUAL — 74-36103-ENG R6
37
Bed exit control system
The patient risk (or fall risk) management functionalities offer a volume range that can be set with the press button
illustrated with the loudspeaker pictogram. The LED indicator by one of the 4 levels will indicate if volume is OFF,
low, medium or high. A range of different tones can be selected with the loudspeaker and treble clef pictogram
press button. The LED indicator will show the selected tone.
The bed exit control system offers different levels of detection: Detection 1 and Detection 2.
To activate/deactivate:
1. Ensure sound level and tone are properly set.
2. Ensure the bed has been properly zeroed.
3. Determine the proper level of detection for bed exit control system.
4. Ensure patient is in bed.
5. Press button for proper level of detection; ensure solid green LED indicator turns ON.
6. Bed exit system is armed.
7. To deactivate bed exit control or stop alarm signal, press button for the selected level of detection for 2 seconds,
ensure LED indicator turns OFF.
The AUTO arm press button offers the possibility to automatically arm the bed exit detection system when patient
gets back in the bed.
To activate/deactivate:
1. Ensure sound level and tone are properly set.
2. Ensure the bed has been properly zeroed.
3. Determine the proper level of detection for bed exit control system.
4. Ensure patient is not in bed.
5. Press the AUTO arm button.
6. Press button for proper level of detection; ensure flashing red LED indicator turns ON.
7. When patient returns in bed, LED indicator on footboard will turn solid green; the bed exit system is armed.
8. To deactivate the Auto arm, press the AUTO button and ensure the LED turns OFF or that the button is no
longer highlighted. To stop an alarm signal, press button for the selected level of detection.
IMPORTANT
The bed exit control settings will not operate if bed is unplugged. In case of power failure, an alarm will be sent to
the nurse call communication system, only if bed exit is armed.