ACCESSORIES
62
USER MANUAL — 74-36103-ENG R6
Folding IV pole 3/4" diameter - FA64507
The folding IV pole FA64507 is equipped of 2 hanging hooks. Each hanging hook has a capacity of 11 lb/5 kg for
a total maximum safe working load of 22 lb/10 kg.
ATTENTION
• IV pole total safe working load is 22 lb/10 kg.
• IV pole is not intended to be used as a push bar; if not avoided permanent damages to the product could
occur.
• Fully extended folding IV pole length is 56", consider maximum bed height of 31" if transporting bed or if
ceiling is low IV pole will move with bed when elevation system activated. Ensure proper clearance.
Installation:
1. Remove screw from bottom of IV pole and
from the IV pole storage support.
2. Insert the folding IV pole in one of both
corner bumpers of head end and insert
the IV pole storage support in the opposite
side 1” socket - as illustrated.
3. Using a # 3 Philips screwdriver, fixe IV pole
and IV pole storage support with screws.
4. Ensure IV pole is properly fixed and folds
properly.
Operation:
1. Ensure IV pole is properly fixed.
2. To adjust the height of the IV pole, hold
the upper section and turn the clamping
ring counterclockwise to loosen the upper
section.
3. Adjust the upper section to required height
and turn the clamping ring clockwise until
upper section is locked.
4. Ensure upper section is properly locked in
place.
5. Store when not in use.
!