OPERATING INSTRUCTIONS
22
USER MANUAL — 200-9011-ENG R1
Battery powered mode: Battery powered mode is activated if power failure occurs or if the bed power cord is
unplugged from wall outlet; it will first fall in sleep mode. Under the battery‑powered
mode, an armed bed exit control system will send a signal to the nurse call
communication system to inform of battery-powered mode; no local sound signal
will be heard. Only the UP/DOWN press buttons, the Trendelenburg and reverse
Trendelenburg press buttons for bed motion and the motion lockout will operate.
Will not operate: 30° backrest positioning and vascular positioning, bed exit control
or scale system, and display screen. The battery charge level is accessible on the
footboard control or on the optional back-up control on frame.
Low level audible signal: Two LED sections of battery pictogram lighted. When a
movement function is activated, functionalities will operate with a low battery level
audible signal.
Recharge audible signal: One red LED section of battery pictogram lighted. When
a movement function is activated, functionalities will not operate and there will be a
constant audible signal.
Bed can be powered down by activating the motion lockouts of the bed from the
footboard and by unplugging the bed power cord from the wall outlet. Ensure wall
outlet is always accessible.