GETTING STARTED
USER MANUAL — 74-36121-ENG R2
25
Zeroing procedure for a bed with display screen with bed exit only
Procedure:
1.
Bed is electrically powered up. Ensure sleep surface is completely flat, siderails are raised, and put the bed
elevation system at intended height of use. Apply the brakes.
2. Ensure bed is not being interfered with or in contact with the wall. Do not
touch bed.
3. Enter the Patient risk management menu from the display screen on the
footboard.
4. Press the ZERO button
EXPORT
Zero
. Enter ID password if required.
5. Follow the instructions; patient is out of bed, install all required equipment,
accessories and mattress.
6. Press the
START
button
Start
on the display screen. Do not touch bed
during procedure.
7. The zeroing is successfully completed when message appears, then screen
returns to Patient risk management menu.
8. If procedure fails, a message appears on screen. Press OK to automatically
return to Zero menu.
9.
Ensure the bed is completely flat and that there is no interference. Start over
the procedure.
Zeroing procedure for a bed with scale option
Cancellation of zeroing process can be done at any time by pressing the “Cancel” button.
Procedure:
1.
Bed is electrically powered up. Ensure sleep surface is completely flat, and put the bed elevation system at
intended height of use. Apply the brakes.
2. From the display screen on the footboard, enter the Scale menu.
3. Press the “zero” button.
4. Password may be required, enter password.
5. Follow the instructions; patient is out of bed, install all required equipments,
accessories and mattress.
6. Press
START
button
Start
on display screen again.
Do not touch bed
while procedure is in progress
.
7.
Zero is successful if a confirmation message appears and the display screen
returns to “Scale” menu.
8. If failure occurs, a message appears to indicate that weight is unstable.
Display screen will stay in Zero menu.
9. Ensure there is no interference and start over the procedure.
Detection 1
Detection 2
Detection 3
Type of Detection
Close
Auto
Arm
Bed Exit Management
November 1st I 12:00
Zero
November 1st I 12:00
Undo Zero
Start
Cancel
Patient must be out of bed.
Put required equipement on
the bed.
Lowered
Cancel
Start
Zero
November 1st 12:00
Undo Zero
Patient shall not
be on bed.
Install equipment.
Zero successfully
completed
OK
Lowered
Cancel
Start
Zero
November 1st 12:00
Undo Zero
Patient shall not
be on bed.
Install equipment.
Zero successfully
completed
OK
Zero
November 1st I 12:00
Undo Zero
Start
Cancel
Patient must be out of bed.
Put required equipement on
the bed.
Scale
November 1st I 12:00
Zero
View Log
Change
Equiment
90.5 kg
Weigh
Max = 272.5 kg
Min = 10 kg
e = 0.5 kg