INTRODUCTION
USER MANUAL — 74-36121-ENG R2
11
EMC information
Guidance and manufacturer’s declaration - emission
The “FL36 Medical Bed” is intended for use in the electromagnetic environment specified below. The customer or user of
the “FL36 Medical Bed” should ensure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment - guidance
RF Emissions
CISPR 11
Group 1
The “FL36 Medical Bed” uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
RF Emissions
CISPR 11
Class B
The “FL36 Medical Bed” is suitable for use in all
establishments,
other than
domestic, and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonics
IEC 61000-3-2
Class A
Flicker
IEC 61000-3-3
Pst: 10 minutes
Plt: 120 minutes
Guidance and manufacturer’s declaration – immunity
The “FL36 Medical Bed” is intended for use in the electromagnetic environment specified below. The customer or user of
the “FL36 Medical Bed” should ensure that it is use in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment -
guidance
ESD
IEC 61000-4-2
±8kV Contact
±15kV Air
±8kV Contact
±15kV Air
Floors should be wood,
concrete or ceramic tile. If
floors are synthetic, the r/h
should be at least 30%
EFT
IEC 61000-4-4
±2kV Mains
±2kV Mains
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
±1kV Differential
±2kV Common
±1kV Differential
±2kV Common
Mains power quality
should be that of a typical
commercial or hospital
environment.
Voltage Dips/Dropout
IEC 61000-4-11
0% Dip for
0.5 Cycle
0% Dip for
1 Cycle
70% Dip for
25 Cycles
0% Dip for
250 Cycles
0% Dip for
0.5 Cycle
0% Dip for
1 Cycle
70% Dip for
25 Cycles
0% Dip for
250 Cycles
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user
of the “FL36 Medical
Bed” requires continued
operation during power
mains interruptions, it
is recommended that
the “FL36 Medical Bed”
be powered from an
uninterruptible power supply
or battery.
Power Frequency
50/60Hz
Magnetic Field
IEC 61000-4-8
30A/m
30A/m
Power frequency magnetic
fields should be that of
a typical commercial or
hospital environment.