OPERATING INSTRUCTIONS
USER MANUAL — 74-36121-ENG R2
45
Footboard control
Footboard control is located on the board at the food end of the bed. Do not use sharp or small pointed objects
on membranes to avoid permanent damages. Pictograms inform the user of positioning possibilities and features.
IMPORTANT
Refer to the
pictograms and definitions
section (page 15) to support identification.
Power up and battery charge level
A green LED indicator constantly ON by the power cord pictogram indicates the bed is power
supplied.
If bed operates on battery, after 2 minutes of non-motion, the sleep mode will activate to save energy; all LED
indicators will turn off, power cord LED indicator will be off. The battery charge level is given by the battery
pictogram with integrated LED.
• No LED section ON, bed is plugged and power supplied: battery is fully charged.
•
3 LED sections flashing green, bed is plugged and power supplied: battery charging,
all functions can operate.
• 3 LED sections solid green, bed is unplugged: battery fully charged - only bed
motions will operate.
• 2 LED sections solid green, bed is unplugged: battery level decreased - only bed
motions will operate, beep at activation.
• 1 LED section solid red, bed is unplugged: battery level is low - none of bed functions
will operate, constant beep signal at activation.
Summary of Contents for OOK Snow FL36
Page 1: ...2 0 2 2 0 3 7 4 3 6 1 2 1 E N G R 2 User manual FL36 FL36 SC MH ...
Page 2: ......
Page 3: ...2 0 2 2 0 3 7 4 3 6 1 2 1 E N G R 2 User manual FL36 FL36 SC MH ...
Page 70: ......
Page 88: ......