OPERATING INSTRUCTIONS
26
USER MANUAL — 74-36105-ENG R5
System signals and messages
The hospital bed FL36 can issue a signal or a message to inform the user that an action is required. It can be
visual and/or audible and/or textual if the bed is equipped of the display screen.
Visual signal
A visual signal may take the form of flashing LEDs on the membrane of the footboard for a bed equipped of a
bed exit system. A reset or a component replacement may be required; contact the maintenance department for
verification.
If the LED for bed power supply, located on the siderails, is flashing when using a one-touch button, it indicates
that the bed must be plugged into the main power supply for usage of this feature.
Audible signal
An audible signal can be caused by an insufficient level of battery charge; plug the bed into the main power supply.
An audible signal can be issued from the control box situated under the seat section. A verification of connections,
a calibration or a component replacement may be required; contact the maintenance department for verification.
IMPORTANT
If the audible signal is emitted from the buzzer situated in the electrical box at head end of the bed or, through
the nurse call system, it indicates that the bed exit system has detected a movement from the patient or, that the
patient has pressed the nurse call button. In both cases, the patient requires assistance from a caregiver.
Textual error message (with display screen option only)
A textual error message can be displayed on the screen located on the footboard. A reset, a verification of
connections, a calibration or a component replacement may be required; contact the maintenance department
for verification.
Transporting bed
Bed can be equipped with casters and braking system, or with floor mounted fixation. Unplug cables from wall
outlets (power cord, nurse call, communication, etc.), store power cord on support brackets and adjust height of
bed for an ergonomic posture before initiating bed transportation. To move bed equipped with casters, release
the brakes and roll the bed to the required location. If transporting bed with a patient in the bed, lock siderails up
and adjust height of bed at the lowest height while as ergonomic as possible. If bed is equipped with the plastic
boards, use the built-in pull and push handles located in the corners of the headboard and footboard. If bed is
equipped of plastic boards without openings, use headboard and footboard to move the bed. Do not use siderails
to move the bed. If not avoided, situation could result in mechanical damages, which could lead to patient fall.
To move a bed equipped of floor mounted fixation, the use of tools will be required for procedure; contact the
maintenance department for assistance. Do not move a bed equipped with floor mounted fixation while a patient
is in the bed.