OPERATING INSTRUCTIONS
USER MANUAL — 74-36105-ENG R5
61
Patient risk management screen menu
Navigate and select:
• Set Detection
to access the bed exit detection
settings. The displayed pictogram informs of
the pre-selected detection level.
• Alarm Settings
to access the volume range
and the tone selection.
• Log
to access the bed exit log.
• Turn Reminder
to set a turn reminder
notification. A nurse call alarm can be
configured to be activated on the bed only or
to be sent to the nurse station.
• Inform
to set up a screen saver configurated to inform caregivers with color coding and/or different
pictograms from the Pictograms library and/or siderail configuration requirements based on patient
condition.
• Close
to exit menu
• Zero
to "zero" the bed exit settings. Menu mentions that patient shall not be in bed and to install equipment
if appropriate prior to start. Bed shall not be touched while zeroing, or cancel command. Screen returns to
the Patient risk management
menu selection once completed. Also possible to select one of the previous
four zeros performed through the Undo Zero screen if the bed is equipped with scale option. Select “apply
zero” to confirm and return to the Patient risk management screen. If bed is not equipped with scale
system, only the last zero performed is available and can be applied. If bed is equipped with scale option,
the function to perform a zeroing or to undo zero is found in the
Scale screen
(page 70).
IMPORTANT
A simplified interface displaying less options is offered for a simplified use. It is possible to change the interface
through the Preference menu, then through Maintenance menu in Bed options menu. On the Standard interface,
the Auto Arm button, the Turn reminder and the Inform button are not displayed. If PIN protected, contact
maintenance for access.
In the Home screen, the Preference box displays the symbol for Lockout and the Inform pictogram if selections
made in Inform. The Preference menu offers short cuts to get to Inform or to Lockouts screen menus.
Depending on configuration selected when ordered, bed’s display presentation on the screen may vary.
Inform
Patient Risk Management
November 1st 12:00
Select
Close
Alarm Settings
Set Detection
Summary of Contents for ooksnow FL36
Page 1: ...2 0 1 7 1 0 7 4 3 6 1 0 5 E N G R 5 User manual FL36 MH...
Page 2: ......
Page 3: ...2 0 1 7 1 0 7 4 3 6 1 0 5 E N G R 5 User manual FL36 MH...
Page 46: ......
Page 58: ......
Page 74: ......
Page 90: ......