14
COMMON OPTIONS AND ACCESSORIES
Hospital Grade Receptacle with Child Safe Cover (852)
1
To access receptacles, insert a screwdriver, key, or
other tool into slot on receptacle cover to remove
child-proof cover.
2
Replace child-proof cover following use.
MULTIPLE SOCKET OUTLET IS TO BE USED FOR MEDICAL EQUIPMENT ONLY (WHERE SAFETY CERTIFICATION
HAS BEEN PERFORMED IN ACCORDANCE TO IEC 60601-1 AND/OR IEC 60601-1-1). USE OF EQUIPMENT NOT
COMPLYING WITH THE EQUIVALENT SAFETY REQUIREMENT OF THIS EQUIPMENT MAY LEAD TO A REDUCED
LEVEL OF SAFETY OF THE RESULTING SYSTEM.
OUTLET IS FOR MEDICAL EQUIPMENT ONLY.
CAUTION: CHECK GROUND CONTINUITY PERIODICALLY.
120V MAXIMUM OUTPUT - 5.0A, 120V
220V MAXIMUM OUTPUT - 2.5A, 220V
WARNING:
DO NOT DISCARD COVER. COVER IS TO BE REAPPLIED AFTER REMOVAL OF EQUIPMENT FROM
MULTIPLE SOCKET OUTLET.
WARNING:
WARNING:
GROUND RELIABILITY CAN ONLY BE ACHIEVED WHEN POWER SUPPLY IS CONNECTED TO AN EQUIVALENT
RECEPTACLE MARKED “HOSPITAL ONLY” OR “HOSPITAL GRADE”.
WARNING:
Service Note:
If the maximum load of the receptacle is exceeded, two circuit breakers will interrupt power. To reset, toggle the far two
circuit breakers located behind the receptacle outlet box in the back of the drawer panel.