SETUP INSTRUCTIONS

Models 4010/4011/5016-650-100 

 

(Power4010, Power4011, Power5016)

To set up your New Power Table:

PLEASE FOLLOW THESE STEPS COMPLETELY

 to ensure 

that your table functions properly PLEASE DO NOT PLUG 

TABLE IN UNTIL PRECEDING STEPS ARE COMPLETED.

1.   Locate the hand control port at the rear of the table base. 
2.   Locate green dot on the port at the rear of the table base. 
3.  Locate green dot on the plug end of the hand control.
4.   Align green dot on plug with green dot on port. 
5.   Plug in hand control completely, ensuring that yellow 

 

‘O’ ring is completely inserted into the port until the 

 

yellow color completely disappears.

 

NOTE: IF YELLOW O-RING IS NOT COMPLETELY 
INSERTED INTO THE PORT, THE TABLE WILL NOT 
FUNCTION. THE TABLE WILL BEEP AND REQUIRE 

 

A CONTROL BOX RESET. THIS IS A SAFETY FEATURE.

6.   Slide the round crew lock forward over the port plug 

 

and turn clockwise to tighten. Ensure completely tight. 

7.   Plug power cord into the port in the exam table. 
8.   Plug power cord into the wall outlet. 

Models 4010/4011/5016-650-200 & 300 

 

(Power4010p, Power4011p, Power5016p, ProGlide4010, 

ProGlide4011, ProGlide5016)

To set up your New Power Table:

PLEASE FOLLOW THESE STEPS COMPLETELY

 to ensure 

 

that your table functions properly PLEASE DO NOT PLUG 

 

TABLE IN UNTIL PRECEDING STEPS ARE COMPLETED.

1.   The table has two (2) ports for the foot control, one (1) 

located at the front and one (1) located at the rear. Locate the 
ports and choose which port to plug in your foot control. 

2.   The table has one (1) port for the hand control located in the 

rear of the table base. 

3.   Start with installing the hand control.
4.   Locate the hand control port at the rear of the table base.
5.   Locate green dot on the port at the rear of the table base. 
6.   Locate green dot on the plug end of the hand control. 
7.   Align green dot on plug with green dot on port. 
8.   Plug in hand control completely, ensuring that yellow 

 

‘O’ ring is completely inserted into the port until the 

 

yellow color completely disappears.

 

NOTE: IF YELLOW O-RING IS NOT COMPLETELY 

 

INSERTED INTO THE PORT, THE TABLE WILL NOT 
FUNCTION. THE TABLE WILL BEEP AND REQUIRE 

 

A CONTROL BOX RESET. THIS IS A SAFETY FEATURE.

9.   Slide the round crew lock forward over the port plug 

 

and turn clockwise to tighten. Ensure completely tight. 

10.  Choose the port in the front or rear of the table for 

 

your foot control. 

11.  Locate green dot on the port.
12.  Locate green dot on the plug end of the foot control.
13.  Align green dot on plug with green dot on port.
14.  Plug in foot control completely, ensuring that yellow 

 

‘O’ ring is completely inserted into the port until the 

 

yellow color completely disappears.

 

NOTE: IF YELLOW O-RING IS NOT COMPLETELY 

 

INSERTED INTO THE PORT, THE TABLE WILL NOT 
FUNCTION. THE TABLE WILL BEEP AND REQUIRE 

 

A CONTROL BOX RESET. THIS IS A SAFETY FEATURE.

15.  Slide the round crew lock forward over the port plug 

 

and turn clockwise to tighten. Ensure completely tight. 

16. Plug power cord into the port in the exam table 
17. Plug power cord into the wall outlet

Part number:  #3-2379LT Rev. A

Summary of Contents for 4010-650

Page 1: ...ACTURE MAXIMUM PATIENT WEIGHT 650 LBS 295 KG OWNER S MANUAL POWER PROCEDURE CHAIR 4010 650 4011 650 5016 650 UMFmedical com Ph 814 266 8726 Fax 814 266 1870 1316 Eisenhower Blvd Johnstown PA 15904 cus...

Page 2: ...ent loading 1 minute on 9 minutes off or 2 minutes on 18 minutes off This product has been evaluated with respect to electrical shock fire mechanical hazards only in accordance with UL60601 1 and CAN...

Page 3: ...60HZ 9AMP Electrical Supply Optional 220V 60HZ 5AMP Power Cord Length 10 FT 3 0m Table Weight 450 lbs 204kg Paper Roll Maximum size 21 0 Long x 3 5 Diameter 53 3 cm x 8 9 cm Caster Base Clearance 0 2...

Page 4: ...cal Supply Standard 120V 60HZ 9AMP Electrical Supply Optional 220V 60HZ 5AMP Power Cord Length 10 FT 3 0m Table Weight 470 lbs 213kg Paper Roll Maximum size 21 0 Long x 3 5 Diameter 53 3 cm x 8 9 cm C...

Page 5: ...cal Supply Standard 120V 60HZ 9AMP Electrical Supply Optional 220V 60HZ 5AMP Power Cord Length 10 FT 3 0m Table Weight 470 lbs 213kg Paper Roll Maximum size 21 0 Long x 3 5 Diameter 53 3 cm x 8 9 cm C...

Page 6: ...n order to prevent misuse when staff is not present Patient Safety Table Lockout To lock and unlock FOR MAXIMUM PATIENT SAFETY LOCK TABLE IMMEDIATELY AFTER EACH USE WARNING Procedure Chair Four Functi...

Page 7: ...Control FEATURES AND OPERATION Hand Foot Control Table Operation DO NOT USE BACKREST AS A SEAT WARNING Height Adjustment Backrest Adjustment Pelvic Tilt Adjustment Legrest Adjustment Home Shown with...

Page 8: ...functions Simultaneous Movement shown as an example To Program 1 Move table to desired position for memory storage 2 Press and hold approximately 3 seconds desired memory position button M1 or M2 in w...

Page 9: ...paper to desired length 1 Pull the stirrups out and unfold 2 Lift slightly and move left or right to position DO NOT SUPPORT THE PATIENT S ENTIRE WEIGHT WITH THE STIRRUPS WARNING DO NOT RAISE LEGREST...

Page 10: ...d Rest 1 Squeeze adjustment button 2 Position the Headrest 3 Release adjustment button to lock headrest in desired place 4 Adjust friction knob so headrest slides smoothly or to lock into position FEA...

Page 11: ...ature 1 Plug into the hand control port 2 Turn the union nut clockwise until fully threaded Back Left to Right 1 Foot control port 2 False Port 3 Hand control port 4 Power on LED 5 Power cord port Fro...

Page 12: ...E IS NOT TO BE USED FOR PATIENT TRANSPORT ENSURE WHEELS ARE IN PATIENT READY POSITION PRIOR TO A PATIENT ACCESSING TABLE TABLE IS NOT TO BE USED FOR PATIENT TRANSPORT CASTERS ARE TO BE USED SOLELY WHE...

Page 13: ...13 Multi Positioning Armrests COMMON OPTIONS AND ACCESSORIES Shown with arm rest accessory...

Page 14: ...MAY LEAD TO A REDUCED LEVEL OF SAFETY OF THE RESULTING SYSTEM OUTLET IS FOR MEDICAL EQUIPMENT ONLY CAUTION CHECK GROUND CONTINUITY PERIODICALLY 120V MAXIMUM OUTPUT 5 0A 120V 220V MAXIMUM OUTPUT 2 5A 2...

Page 15: ...otrol BirexSE Bleach 1 5 20 bleach Bleach 1 9 10 bleach Bleach Rite Disinfecting Spray CaviCide CaviCide1 CaviCide AF Diversey Accel INTERVention Wipes Diversey Avert Sporicidal Disinfectant Cleaner E...

Page 16: ...Facilities for bleach alternative cleaners Note Immediately remove any fluid spilled on upholstery surface Antimicrobial UMF Medical Upholstery providers outstanding protection in difficult medical a...

Page 17: ...in whole or in part by mishandling misuse abuse negligence alteration accident freight damage tampering or failure to seek and obtain repair or replacement in a timely manner Products which are not i...

Page 18: ...the ports and choose which port to plug in your foot control 2 The table has one 1 port for the hand control located in the rear of the table base 3 Start with installing the hand control 4 Locate th...

Reviews: