Perform a measuring campaign
49/92
4. The scale should be placed on a vibration-free work table. The
work table should not be used for other purposes during a
HYPROP measurement.
5. During a single mode measurement we strictly recommend to fix
the cable of the sensor unit.
6. The scale must be levelled out. Most scales have a bubble level.
7. Set the energy manager of your laptop to non-stop operation.
Open the energy option manager and set „Power-down“ and
„Stand-by“ to „Never“. If the laptop powers down or goes to the
stand-by mode no readings are stored.