31
9.3 GLP-compliant print
At calibration, the HYPROP BALANCE Series enables printing in a format compliant with GLP. Set
the function to the GLP print support ([0.GLP 1]) side, then connect the balance to a printer or other
equipment. When calibration is complete, the balance prints out the format illustrated below.
Since only items are printed, the data should be entered manually by the person who conducted
calibration.
※
Reference:
“4.2 Description of Functions” (Page 13)
“9.2 Calibration of HYPROP BALANCE Series” (Page 29)
GLP-Compliant Print Form
**CALI BRATI ON**
MODEL:
S/N:
I D:
DATA:
TI ME:
*CAL. END
NAME:
***************
←見出し
←はかり の型名
←はかり の製造番号
←I D番号
←校正し た 日付
←校正し た 時刻
←校正終了
←担当者署名
Title
Model of balance
Serial No. of balance
ID No.
Date of calibration
Time of calibration
End of calibration
Signature of person in charge
**CALI BRATI ON**
MODEL:
S/N:
I D:
DATE:
TI ME:
*CAL. END
NAME:
***************
※
To operate with CSP-160 printer, set On in DIP switch 3 (Printing controlled by scale), and OFF in
DIP switch 7(disable Manual printing).
※
For more information on the printer, refer to the printer
’s operation manual.
Caution:
Printing is done only when calibration ends normally. The balance will not print if calibration ends
abnormally with [3-Err], [4-Err], or [StOP] being issued.