Programming Manual MSO/UPO2000 Series
103 / 156
Instruments.uni-trend.com
:SBUS:LIN:QUALifier {SYNC|ID|DATA|IDDATA|WAKE|SLEEP|ERROR}
:SBUS:LIN:QUALifier?
Functional description
:
This command is used to set LIN bus trigger condition of the oscilloscope.
{SYNC|ID|DATA|IDDATA|WAKE|SLEEP|ERROR}: synchronization, identifier, data, ID and data, wake frame,
sleep frame, error.
Return format
:
Query returns {SYNC|ID|DATA|IDDATA|WAKE|SLEEP|ERROR}.
For example
:
:SBUS:LIN:QUALifier SYNC
Set LIN bus trigger condition to SYNC.
:SBUS:LIN:QUALifier?
Query returns SYNC’
:SBUS:LIN:ID
Command format
:
:SBUS:LIN:ID <string>
:SBUS:LIN:ID?
Functional description
:
This command is used to set LIN bus indetifier data of the oscilloscope. Binary character data presented
by parameter 0, 1 or X. X represents uncertainty. Its range is 0x0~0Xff.
Return format
:
Query returns binary character string data.
For example
:
:SBUS:LIN:ID "X00X00X1"
Set LIN bus indetifier data to X00X00X1.
:SBUS:LIN:ID?
Query returns X00X00X1.
:SBUS:
LIN
:TRIGger:DATA:LENGth
Command format
:
:SBUS
:
LIN
:TRIGger
:DATA:
LENGth
<length>
:SBUS
:
LIN
:TRIGger
:DATA:
LENGth
?
Functional description
:
This command is used to set LIN bus trigger data length of the oscilloscope. It can take value from 1~8.
Return format
:
Query returns LIN bus trigger data length of the oscilloscope, it is integer data.
For example
:
:SBUS
:
LIN
:TRIGger
:DATA:
LENGth
2
Set LIN bus trigger data length to 2 bytes.
:SBUS
:
LIN
:TRIGger
:DATA:
LENGth
?
Query returns 2.
:SBUS:LIN:DATA
Command format
:
:SBUS:LIN:DATA <string>