Programming Manual MSO/UPO2000 Series
115 / 156
Instruments.uni-trend.com
AWG:CHANnel<n>:BASE:OFFSet
Command format
:
AWG:CHANnel<n>:BASE:OFFSet <voltage>
AWG:CHANnel<n>:BASE:OFFSet?
Functional description
:
This command is used to set DC offset of the specified channel.
<voltage> represents voltage, unit is V. The range is 0~± the maximum DC under the current load.
the maximum DC under the current load = the current load *10/(50+ the current load) – the current AC
minimum value /2;
AC minimum value is 2mVpp, take 0 in DC mode.
<n>: channel number, n take value from 1, 2.
Return format
:
Query returns DC offset of the specified channel in scientific notation.
For example
:
AWG:CHANnel1:BASE:OFFSet 2
Set DC offset of channel 1 to 2 V.
AWG:CHANnel1:BASE:OFFSet?
Query returns 2e+0.
AWG:CHANnel<n>:BASE:HIGH
Command format
:
AWG:CHANnel<n>:BASE:HIGH <voltage>
AWG:CHANnel<n>:BASE:HIGH?
Functional description
:
This command is used to set output high value of the specified channel signal.
<voltage> represents voltage, unit is assigned by the current channel.
<n>: channel number, n take value from 1, 2.
Return format
:
Query returns output high value of the specified channel signal in scientific notation.
For example
:
AWG:CHANnel1:BASE:HIGH 2
Set output high value of channel 1 signal to 2 V.
AWG:CHANnel1:BASE:HIGH?
Query returns 2e+0.
AWG:CHANnel<n>:BASE:LOW
Command format
:
AWG:CHANnel<n>:BASE:LOW <voltage>
AWG:CHANnel<n>:BASE:LOW?
Functional description
:
This command is used to set output low level of the specified channel signal.