Programming Manual MSO/UPO2000 Series
128 / 156
Instruments.uni-trend.com
Query returns 1 or 0, it respectively represents ON or OFF.
For example
:
:LA:BUS1
:DISPlay ON
Turn on LA bus 1 decoding status to display decoding waveform.
:LA:BUS1
:DISPlay?
Query returns 1.
:LA:BUS<n>:BASE
Command format
:
:LA:BUS<n>
:BASE {ASCII | BINary | HEX | DEC| WAV}
:LA:BUS<n>
:BASE?
Functional description
:
This command is used to set LA bus display format of the oscilloscope.
<n>: integer data {1|2}, it respectively represents bus 1~2.
Return format
:
Query returns {ASCII | BINary | HEX | DEC | WAV }.
For example
:
:LA:BUS1
:BASE BIN
Set LA bus 1 decoding display in binary format.
:LA:BUS1
:BASE?
Query returns BINary.
:LA:BUS<n>:EVENt
Command format
:
:LA:BUS<n>
:EVENt { {1|ON} | {0|OFF} }
:LA:BUS<n>
:EVENt?
Functional description
:
This command is used to turn on/off LA bus event list.
<n>: integer data {1|2}, it respectively represents bus 1~2.
Return format
:
Query returns 1 or 0, it respectively represents ON or OFF.
For example
:
:LA:BUS1
:EVENt ON
Turn on LA bus 1 event list.
:LA:BUS1
:EVENt?
Query returns 1.
:LA:BUS<n>:DATA?
Command format
:
:LA:BUS<n>:DATA?
Functional description
:
This command is used to read LA decoding event list data of the oscilloscope.
Return format
:
Query returns LA decoding event list data, returned data is comform with
This command is used to