Programming Manual MSO/UPO2000 Series
45 / 156
Instruments.uni-trend.com
This command is used to set video standard.
Return format
:
Query returns {
NTSC | PAL | SECAM }.
For example
:
:TRIGger:VIDeo:STANdard NTSC
Set video standard to NTSC.
:TRIGger:VIDeo:STANdard?
Query returns NTSC.
:TRIGger:VIDEO:LINE
Command format
:
:TRIGger:VIDEO:LINE <value>
:TRIGger:VIDEO:LINE?
Functional description
:
This command is used to set the assigned line. <value> represents the assigned line, range is relative to
video standard.
Return format
:
Query returns the current assigned line.
For example
:
:TRIG:VIDEO:LINE 50
Set the assigned line of video synchronization to 50.
:TRIG:VIDEO:LINE?
Query returns 50.
Slope Trigger
:TRIGger:SLOPe:QUALifier
Command format
:
:TRIGger:SLOPe:QUALifier {GREaterthan | LESSthan | INRange}
:TRIGger:SLOPe:QUALifier?
Functional description
:
This command is used to set the setting condition of slop time, which includes GREaterthan (greater than),
LESSthan (less than) and INRange (between).
Return format
:
Query returns {GREaterthan | LESSthan | INRange}.
For example
:
:TRIGger:SLOPe:QUALifier GRE
Set slop condition to GREaterthan.
:TRIGger:SLOPe:QUALifier?
Query returns GREaterthan.
:TRIGger:SLOPe:SLOPe
Command format
:
:TRIGger:SLOPe:SLOPe {POSitive|NEGative}
:TRIGger:SLOPe:SLOPe?
Functional description
: