Programming Manual MSO/UPO2000 Series
8 / 156
Instruments.uni-trend.com
:SYSTem:CAL
Command format
:
:SYSTem:
CAL
Functional description
:
This command is used to set the self-calibration of the system, it cannot be normal communicated during
self-calibration.
:SYSTem:CLEAr
Command format
:
:SYSTem:
CLEAr
Functional description
:
Empty all the saved waveforms and data settings.
:SYSTem:CYMOmeter
Command format
:
:SYSTem:CYMOmeter {1 | ON} | {0 | OFF}
:SYSTem:CYMOmeter?
Functional description
:
This command is used to turn on/off the frequency meter.
Return format
:
Query returns the status of frequency meter, 1 represets ON, 0 represents OFF.
For example
:
:SYSTem:CYMOmeter ON
Turn on frequency meter.
:SYSTem:CYMOmeter?
Query returns 1.
:SYSTem:CYMOmeter:FREQuency?
Command format
:
:SYSTem:CYMOmeter:FREQuency?
Functional description
:
This command is used to acquire the frequency value of the frequency meter measurement.
Return format
:
Query returns the frequency value of the frequency meter measurement, and invalid value returns *.
For example
:
:SYSTem:CYMOmeter:FREQuency? Query returns 1.3.
:SYSTem:SQUare:SELect
Command format
:
:SYSTem:SQUare:SELect { 10 Hz | 100 Hz | 1 kHz | 10 kHz }
:SYSTem:SQUare:SELect?
Functional description
:
This command is used to select square wave output.