13.1 System Function Setting
(1) Automatic Calibration
Automatic calibration allows the oscilloscope to work optimally to obtain the most accurate
measurements. This function can perform at any time, especially when the ambient temperature
range reaches or exceeds 5
℃
. Before performing the automatic calibration operation, make sure
that the oscilloscope is operating for more than 20 minutes.
(2) System Information
System Information can view the system information of the oscilloscope, which includes
manufacturer, model number, SN code, software version number, logical version number, hardware
version number, website, website user name & password and date.
(3) Language
Language can set system language, English, simplified Chinese. By default, language display is the
current setting.
(4) Square Waveform Output
Square Output can set the output frequency of local square wave: 10Hz, 100Hz, 1kHz, 10kHz.1kHz
by default.
(5) Frequency Meter
Frequency Meter can set the state of frequency meter (on/off), when the frequency meter is enabled,
the frequency information will displayed on the top of screen. Frequency meter is the counter of
trigger event frequency in the trigger channel, display as hardware 7 digitals frequency meter.
(6) Output Selection
Output Selection can select output signal by AUX OUT port
, it can select “trigger” or “pass/fail”.
Trigger: AUX OUT port outputs the trigger synchronization signal.
Pass/Fail: AUX OUT port outputs the pass/fail signal.
The default selection is “trigger”.
(7) Delete Data
Delete Data can clear the stored waveform setting file, and etc.
(8) Internet Setting
When the device is connected with available internet, IP setting can set the IP, subnet mask and
gateway of the oscilloscope.
a. Mode
:
manual or auto
Manual
:
it can set IP address and subnet maskby by manual
Auto
:
Only can view the IP address and submask