12.3 Pass/Fail
The Pass/Fail uses a template to detect whether the input signal satisfies the template requirements. If the input signal
exceeds the limited range of the template, it is judged as failed.
(1) Function Introduction
Press UTILITY PgDn pass/fail to enter its menu.
Enable:
Press enable, select on to make the relevant settings for the Pass/Fail.
Output:
Press output to select fail or pass.
①
Fail: Set the AUX interface on the back panel of the oscilloscope to output pulses when “fail” and produce beeps.
②
Pass: Set the AUX interface on the back panel of the oscilloscope to output pulses when “pass” and produce beeps.
Note:
AUX output selection needs to be switched to Pass/Fail.
Source:
Press source, select CH1, CH2, CH3, or CH4 to set the test source of Pass/Fail.
Message:
Press message, select on and the test result statistics will be displayed at the upper left corner of the screen, as shown below:
Total_wfs means the total number of frames tested; Pass_wfs means the number of frames passed; Fail_wfs means the
number of frames failed.
Stop Setting:
Press PgDn
stop setting to enter its menu as follows:
Descriptions
Set the Pass/Fail function to stop the test automatically after
it reaches the specified threshold value of pass
Set the Pass/Fail function to stop the test automatically after
it reaches the specified threshold value of failure
Set the stop condition
Use the Multipurpose knob to set the stop condition threshold
Returns to previous menu, the Pass/Fail menu
Options
Pass Count
Fail Count
>=, <=
Functions
Stop Type
Condition
Threshold
Return
Template Setting:
Press PgDn template setting to enter its menu. As shown below:
Descriptions
Select a channel waveform in CH1~CH4 and plus the horizontal
and vertical tolerance as the conditions for creating a template
Use the Multipurpose knob to select the horizontal tolerance
Use the Multipurpose knob to select the vertical tolerance
Creates a template with above conditions
Returns to previous menu, the Pass/Fail menu
Functions
Reference Waveform
Horizontal Tolerance
Vertical Tolerance
Create Template
Return
Options
Ch1, CH2,
CH3, CH4
1 ~ 255
1 ~ 255
55