14
3. Input Current Jack: There is a 5x20mm fuse
inside the “A” input jack. Over range
measurement will burn the fuse and replacement
should use the fuse of the same specification.
“10A” input jack does not have fuse protection.
4. The meter will show the room temperature if
the temperature probe is not connect to the
object under testing. The meter will only shows
the object temperature if the probe is connect
to it.
5. When the meter is not in use for more than
15 minutes, the power will be automatically
turned off. Just release & press the “POWER”
button, the meter will be turned on again.
6. Battery or fuse replacement should only be
done after the test leads have been disconnected
and power is off, to open the battery door,
see the following diagram.
Model UT2000: OPERATING MANUAL