l
Maximum Voltage between any Terminals and grounding: Refer to different
range input protection voltage.
l
Display: 3 3/4 digits LCD display, Maximum display 4000.
l
Polarity: Auto
l
Overloading: Display
OL
or
–OL
.
l
Battery Deficiency: Display
.
l
Sampling: 3 times per second.
l
Measurement Deviation: The conductor being meaured is not placed in the
center of the jaw during AC/DC current
measurement, it will cause
extra ±1%
deviation based on the stated accuracy.
l
Drop Test: 1 meter drop test passed.
l
Max. Jaw Size: 55mm diameter.
l
Projected Max. Current conductor size: 45mm diameter.
l
Electro-Magnetic: When carrying out measurement near the electro-magnetic,
it may cause unstable or wrong reading.
l
Power: 1 x 9V battery (6LF22 1604A)
Specifications
A.General Specifications:
Model UT209: OPERATING MANUAL
33