Measurement Operation (9)
22
Model UT33A: OPERATING MANUAL
Note
l
The test leads can add 0.1
Ω
to 0.3
Ω
of error to resistance measurement. To
obtain precision readings in low-resistance measurement, that is the range of
200
Ω
, short-circuit the input terminals beforehand and record the reading
obtained (called this reading as X). (X) Is the additional resistance from the test
lead. Then use the equation: measured resistance value (Y) - (X) = precision
readings of resistance.
l
For high-resistance measurement (>1M
Ω
), it is normal taking several seconds
to obtain a stable reading.
l
When resistance measurement has been completed, disconnect the connection
between the testing leads and the circuit under test.