Measurement Operation (4)
Note
l
If the value of voltage to be measured is unknown, use the maximum measurement
position (500V) and reduce the range step by step until a satisfactory reading
is obtained.
l
The LCD displays
“1”
indicating the existing selected range is overload, it is
required to select a higher range in order to obtain a correct reading.
l
In each range, the Meter has an input impedance of approx. 10M
Ω
. This loading
effect can cause measurement errors in high impedance circuits. If the circuit
impedance is less than or equal to 10k
Ω
, the error is negligible (0.1% or less).
l
When AC voltage measurement has been completed, disconnect the connection
between the testing leads and the circuit under test.
16
Model UT33B/C/D: OPERATING MANUAL