UT3500 series
User Manual
17
Figure 3-2-1 A UT-L80 Battery Internal Resistance Test Pen-Crown Probe
(Standard Accessory)
Figure 3-2-1 B UT-L82 Battery Internal Resistance Kelvin Test Lead- Nipper Clip
(Standard Accessory)
Figure 3-2-1 C UT-L81 Battery Internal Resistance Test Lead- Alligator Clip
(Optional Accessory)
3.2.2 Connection of Test Lead
From the perspective of safety and test accuracy, it is recommended to use the standard
test lead for measurement. Before measurement, follow the steps below to connect the
test lead to the test terminals of the instrument.
1. Before connecting the test line, please make sure that the power switch of the
instrument is OFF.
2. Make sure there is no connection at the top of the test leads of the four terminals.
3. Connect the test leads of the four terminals to the input terminals of the instrument
according to the method shown in Figure 3-2-2. The specific operations are as
follows:
Connect the black test lead to the black terminal hole of the instrument and the red
test lead to the red terminal hole.
The
▲
symbol of the black test lead must match the black SENSE hole on the front
panel of the instrument, and the
▲
symbol of the red test lead must match the red
SENSE hole.
Red
Black
Red Crown Probe Test Pen
Black Crown Probe Test Pen
Red
Black
Red Nipper Clip
Black Nipper Clip
Red
Black
Red Alligator Clip
Black Alligator Clip